MICHIGAN DEPARTMENT OF ENVIRONMENTAL QUALITY
AIR QUALITY DIVISION
As Amended
R 336.2001 Performance tests by owner.
Rule 1001. (1) The department may require the owner or
operator of any source of air contaminant to conduct acceptable performance
tests, at the owner's or operator's expense, in accordance with R 336.2003
under any 1 of the following conditions:
(a) Prior to issuance of a permit to operate.
(b) The source is determined to be in violation
of R 336.1301 and the potential emissions exceed 100 tons per year.
(c) The owner or operator of the source has not
submitted an acceptable performance test, in accordance with R 336.2003, that
demonstrates that the source is in compliance with the department’s rules and
with the conditions specified in the permit to install.
(d) The source of air contaminant is located in
an area designated as nonattainment for 1 or more air pollutants, and more than
12 months have expired since the date of the last performance test for such
designated nonattainment pollutants.
(e) The source of air contaminant has potential
emissions in excess of 100 tons per year, is located in an area designated as
attainment for 1 or more air pollutants, and more than 36 months have expired
since the date of the last performance test for such designated attainment
pollutants.
(f) After completion of a compliance program.
(2) Performance tests required by subrule (1) of
this rule shall be conducted
within 60 days following receipt of written notification from the department,
unless otherwise authorized by the department.
(3) Not less than 7 days before performance tests
are conducted, the owner of a source of air contaminant, or his or her
authorized agent, shall notify the department, in writing, of the time and
place of the performance tests and who shall conduct them. A representative of the department shall have
the opportunity to witness these tests.
(4) Results of performance tests shall be
submitted to the department in the format prescribed by the applicable
reference test method within 60 days after the last date of the test.
History: 1979 ACS 1, Eff.
R 336.2002 Performance tests by department.
Rule 1002. (1)
The department may conduct performance tests in accordance with
R 336.2003 at any source of air contaminant, on behalf of the state, at a
reasonable time and at the state's expense.
During the conduct of such tests, the department may obtain samples of
any air contaminant and samples of any material entering or exiting the source
or air-cleaning device for the purpose of evaluating pollutant emissions with
respect to process operating conditions.
(2) The department shall provide written
notification to the owner or operator of a source of the department’s intent to
conduct performance tests pursuant to subrule (1). Within 30 days of receipt of such notification,
the owner or operator shall provide, and bear the expense of, performance test
facilities as specified by the department, including the following:
(a) Sampling ports adequate for reference test
methods applicable to the source.
(b) Safe sampling platforms as required.
(c) Safe access to sampling platforms.
(d) A suitable power source within 50 feet of any
sampling location designated by the department.
Upon request,
additional time for installing the required performance test facilities may be
authorized by the department for special situations.
(3) The owner shall not be responsible for
providing sampling instruments and sensing devices.
(4) Results of performance tests shall be
furnished to the owner or operator, or both, in the format prescribed by the
applicable reference test method within 60 days following the last date of the
test.
History: 1979 ACS 1, Eff.
R 336.2003 Performance test criteria.
Rule
1003. (1) Performance tests shall be conducted and data
reduced according to the reference test methods listed in R 336.2004, unless the department does
any of the following:
(a) Specifies or approves, in specific cases, the
use of a reference test method with minor changes in procedures or equipment.
(b) Approves the use of an equivalent method.
(c) Specifies or approves the use of an
alternative method if an applicable reference test method does not exist for a
specific air contaminant or source of air contaminant.
(2) A performance test shall consist of a minimum
of 3 separate samples of a specific air contaminant conducted within a 36-hour
period, unless otherwise authorized by the department. Each of the 3 separate samples shall be
obtained while the source is operating at a similar production level. For the purpose of determining compliance
with an applicable emission limit, rule, or permit condition, the arithmetic
mean of results of the 3 samples shall apply.
If a sample is accidentally lost or conditions occur in which 1 of the
3 samples must be discontinued because of forced shutdown, failure of an
irreplaceable portion of the sampling train, extreme meteorological conditions,
or other circumstances beyond the owner's or operator's control, then
compliance may, upon the approval of the department, be determined using the
arithmetic mean of the results of 2 samples.
(3) All performance tests shall be conducted
while the source of air contaminant is operating at maximum routine operating
conditions, or under such other conditions, within the capacity of the
equipment, as may be requested by the department. Other conditions may include source operating
periods of start-up, shutdown, or such other operations, excluding malfunction,
specific to certain sources. Routine
operating conditions shall also include those specified within a permit to
install or a permit to operate. The
owner or operator shall make available to the department such records as may be
necessary to determine the conditions of source operation that occurred during
the period of time of the performance test.
(4) For any source that is subject to an emission
limitation calculated to 50% excess air,
the multipoint, integrated sampling procedure of method 3 shall be used
for gas analysis. For all other sources
that require a determination of the molecular weight of the exhaust, any
optional sampling procedure of method 3 may be used. Alternatives or modifications to procedures
are subject to the approval of the department.
(5) For reference test methods 5B and 5C, the
minimum volume per sample shall be 30 cubic feet of dry gas corrected to
standard conditions (70 degrees Fahrenheit, 29.92 in. Hg.). Minimum sample time shall be 60 minutes,
which may be continuous or a combination of shorter sampling periods for
sources that operate in a cyclic manner.
Smaller sampling times or sample volumes, when necessitated by process
variables or other factors, may be approved by the department.
History: 1979 ACS 1, Eff.
R 336.2004 Appendix A; reference test methods; adoption of federal reference test methods.
Rule 1004. (1) The following federal reference test methods,
described in the provisions of 40 C.F.R. part 60, appendix A (2005), are the reference test
methods for performance tests required pursuant to the provisions of this part:
(a) Method 1 - Sample and velocity traverse for
stationary sources.
(b) Method 1A - Sample and velocity traverses for
stationary sources with small stacks or ducts.
(c) Method 2 - Determination of stack gas
velocity and volumetric flow rate (type-S pitot tube).
(d) Method 2A - Direct measurement of gas volume
through pipes and small ducts.
(e) Method 2C - Determination of stack gas
velocity and volumetric flow rate in small stacks and ducts (standard pitot tube).
(f) Method 2D - Measurement of gas volumetric
flow rates in small pipes and ducts.
(g) Method 3 - Gas analysis for the determination
of dry molecular weight.
(h) Method 4 - Determination of moisture content
in stack gases.
(i) Method 6 - Determination of sulfur dioxide
emissions from stationary sources.
(j) Method 7 - Determination of nitrogen oxide
emissions from stationary sources.
(k) Method 8 - Determination of sulfuric acid
mist and sulfur dioxide emissions from stationary sources.
(l) Method 9 - Visual determination of the
opacity of emissions from stationary sources.
(m) Method 10 - Determination of carbon monoxide
emissions from stationary sources.
(n) Method 10B - Determination of carbon monoxide
emissions from stationary sources.
(o) Method 18 - Measurement of gaseous organic
compound emissions by gas chromatography.
(p) Method 21 - Determination of volatile organic
compound leaks.
(q) Method 24 - Determination of volatile matter
content, water content, density, volume solids and weight solids of surface
coatings.
(r) Method 24A - Determination of volatile matter
content and density of printing inks and related coatings.
(s) Method 25 - Determination of total gaseous
nonmethane organic emissions as carbon.
(t) Method 25A - Determination of total gaseous
organic concentration using a flame ionization analyzer.
(u) Method 27 – Determination of vapor tightness
of gasoline delivery tank using pressure-vacuum test.
(2)
The reference test methods listed in
subrule (1) of this rule are adopted by reference in this rule. Copies of the test methods may be inspected
at the
(a) Visible emissions from a scarfing operation
at a steel manufacturing facility shall be determined as specified in reference
test method 9A, which is described in R 336.2030.
(b) Visible emissions from a coke oven pushing
operation and fugitive coke oven visible emissions shall be determined as
specified in reference test method 9B, which is described in R 336.2031.
(c) Visible emissions, fugitive and nonfugitive,
from basic oxygen furnace operations, hot metal transfer operations, and hot
metal desulfurization operations shall be determined as specified in reference
method 9C, which is described in R 336.2032.
(a) Reference test method 5B, which is described
in R 336.2011.
(b) Reference test method 5C, which is described
in R 336.2012.
(c) Reference test method 5D, which is described
in R 336.2013.
(d) Reference test method 5E, which is described
in R 336.2014.
(e) "Standard Methods for the Examination of
Water and Wastewater," (14th edition), section 208C, as described and
modified in R 336.2033.
History: 1979 ACS 1, Eff. Jan. 19, 1980; 1985 MR 2, Eff. Feb. 22, 1985; 1989 MR 4, Eff. Apr. 19, 1989; 1993 MR 4, Eff. Apr. 28, 1993; 2000 MR 18, Eff. Nov. 30, 2000; 2002 MR 5, Eff. Mar. 19, 2002; 2006 MR 4, Eff. Feb. 22, 2006.
R 336.2005 Reference test methods for state-requested tests of delivery
vessels.
Rule
1005. The following reference
test method shall be used to detect gasoline vapor leaks by a combustible gas
detector:
(a) Principle.
A combustible gas detector is used to indicate any incidence of leakage
from gasoline delivery vessel tanks and vapor control systems. This qualitative monitoring procedure is an
enforcement tool to confirm the continuing existence of leak-tight conditions.
(b) Applicability. This method is applicable to determining
leak-tightness of gasoline delivery vessel tanks during loading without taking
the delivery vessel tank out of service. The method is applicable only if the
vapor control system does not create back pressure in excess of the pressure
limits of the delivery vessel tank compliance leak test. For vapor control systems, this method is
applicable to determining leak-tightness at any time.
(c) Apparatus and specifications. The following apparatus shall be used:
(i) Manometer.
Liquid manometer, or equivalent, capable of measuring up to 0.9 pounds
per square inch (24.9 inches of water) gauge pressure within 0.003 pounds per
square inch (0.1 inches of water) precision.
(ii) Combustible gas detector. A portable hydrocarbon gas analyzer with
associated sampling line and probe which complies with all of the following
provisions:
(A) Safety.
The device is certified as safe for operation in explosive atmospheres.
(B) Range.
The device shall have a minimum range of 0 to 100% of the lower explosive
limit (LEL) as propane.
(C) Probe diameter. The sampling probe shall have an internal
diameter of 0.625 centimeters (1/4 inch).
(D) Probe length.
The probe sampling line shall be of sufficient length for easy
maneuverability during testing.
(E) Response time. The response time for full-scale deflection
shall be less than 8 seconds for a detector with a sampling line and probe
attached.
(d) Test procedure. The following test procedure shall be
complied with:
(i) Pressure.
Place a pressure tap in the terminal, plant, or service station vapor
control system as close as possible to the connection with the delivery vessel
tank. Record the pressure periodically
during testing.
(ii) Calibration.
Calibrate the combustible gas detector with 2.2% propane, by volume, in air for 100% lower explosive limit response.
(iii) Monitoring procedure. During loading or unloading, check the
periphery of all potential sources of leakage of the delivery vessel tank and
of the terminal, plant, or service station vapor collection system with a
combustible gas detector. The check
shall comply with the following procedure:
(A) Probe distance. The probe inlet shall be 2.5 centimeters from
the potential leak source.
(B) Probe movement. Move the probe slowly (2.0 centimeters per
second). If there is any meter
deflection at a potential leak source, move the probe to locate the point of
highest meter response.
(C) Probe position. As much as possible, the probe inlet shall be
positioned in the path of (parallel to) the vapor flow from a leak.
(D) Wind. Attempt, as much as possible, to block the wind from the area being
monitored.
(iv) Recording. Record the highest detector reading and
location for each incidence of leakage.
History: 1979 ACS 7, Eff. Aug. 22, 1981; 1989 MR 4, Eff. Apr. 20, 1989; 2002 MR 5, Eff. Mar. 19, 2002; 2006 MR 4, Eff. Feb. 22, 2006.
R 336.2006 Reference test method serving as alternate version of federal reference test method 25 by incorporating Byron analysis.
Rule 1006. (1) When using the alternate version of federal reference test method 25 incorporating the Byron analysis, the procedures in method 25, which are described in R 336.2004, shall be followed, except that all of the following parts in method 25 are amended to read as follows:
1.2 Principle. An emission sample is withdrawn from a stack
at a constant rate through a stainless steel absorber tube packed with porasil;
the gaseous portion of the sample is pulled past a battery-operated sampling pump
into a tedlar bag. After sampling is
complete, the contents of the tedlar bag are analyzed on an automated gas
chromatograph (GC), and the sample in the porasil packed tube is heated to
remove all components for analysis on the GC. The GC separates CO, CO2,
and CH4 from the nonmethane organics (NMO), then
converts the NMOs to methane for analysis.
2. Apparatus. The sampling system consists of a nonmethane
organic (NMO) absorber tube, a sampling pump, and a sample bag (figure
25-1). The analytical system has 2
parts--the oven for removing the sample from the absorber tube and an automated
gas chromatograph (GC).
2.1 Sampling. All of the following equipment is required,
as shown in figure 25-1:
2.1.1 Heated
probe. 6.4-millimeter (mm) (1/4-inch
(in.)) outside diameter (o.d.) stainless steel tubing with a heating system
that is capable of maintaining a gas temperature at the exit end of not less
than 129 degrees Centigrade (265 degrees Fahrenheit). The probe shall be equipped with a thermocouple
at the exit end to monitor the gas temperature.
The nozzle is an elbow fitting that is attached to the front end of the
probe while the thermocouple is inserted in the side arm of a tee fitting that
is attached to the rear of the probe.
The probe is wrapped with a suitable length of high-temperature heating
tape and then covered with 2 layers of glass cloth insulation and 1 layer of
aluminum foil.
2.1.2 Heated
prefilter--only for stacks with possible particulate matter interference. A stainless steel filter
holder with a 47-mm type A/E fiberglass filter without organic
binder. The entire prefilter shall be
maintained at 110 degrees Celsius. Note
- if it is not possible to use a heating system for safety reasons, an unheated
system with an instack filter is a suitable alternative.
2.1.3 NMO
absorber tube. 1/2-inch inside diameter
(i.d.) stainless steel tube packed with porasil (thermally stable silica gel).
2.1.4 1/4-inch
o.d. teflon line that is 2 to 4 feet long.
2.1.5 Battery-operated
diaphragm sampling pump with kurz digital mass flow meter. Total flow is integrated electronically to
measure flow with an accuracy of 1% at any flow rate. (Byron instruments model 90).
2.1.6 Sample
bag. 0.3 mil-tedlar,
1/2-cubic foot capacity. The
sample bag undergoes nitrogen purge cycle until analysis exhibits zero carbon
content in the sample bag.
2.2 Analysis. The following equipment is required:
2.2.1 Sample
recovery on the adsorber tube is done in a Byron model 75 oven in 2 stages,
each stage requiring a 0.3-mil tedlar bag that has a 1/2-cubic foot capacity.
2.2.2 Analysis
is done on a Byron model 401 gas chromatograph (GC) that meets all criteria
specified in method 25, section 2.2.2.
2.3 NMO
analyzer. The NMO analyzer is a Byron
model 401 gas chromatograph (GC).
(Remainder of 2.3 as stated in method 25)
2.3.5.2 Range. A full scale range of 1 to 10,000 parts per
million (ppm) CH4. Signal
attenuators shall be available to produce a minimum signal response of 10% of
full scale.
3.1.1 Delete
(dry ice is not required).
4.1.1 Model
90 and model 75 flow meter calibration. The model 90 sample pump inlet is attached to
the outlet of the model 75 oven. Air is
passed through the system at the rate that will be used in sampling and for the
total volume anticipated to be sampled.
If the flow meters on the 2 instruments do not agree within 0.01 liters,
then adjust the meter on the model 90 until agreement is within 0.01
liters. After making any correction, run
a full calibration again.
4.1.2 Sample
train assembly. Assemble the probe
(prefilter if needed), adsorber tube, and teflon line
to the inlet of the model 90. Attach a
short (8 to 10 inches) flexible line to the outlet of model 90. Have a completely clean evacuated tedlar
sample bag nearby for collection of sample to be analyzed.
4.1.3 Pretest
leak check. Stopper the inlet of the
probe and place the flexible tube on the outlet of the model 90 in a small open
container of water. Turn on the sampling
pump. For a satisfactory leak check,
bubbling should cease within 1 minute.
If the leak check is unsatisfactory, tighten the fittings or change
parts until a satisfactory leak check is obtained.
4.1.4 Sampling
train operation. Place the probe and the
front portion of the adsorption tube in the stack. If the stack has a temperature higher than
ambient, allow time for the probe to heat before starting the sample pump. Start the model 90 pump and adjust to the
desired flow, usually about 90 ml/min.
After about 0.1 liter of sampling, or equivalent to the volume of air
that is displaced in the sample system before the flowmeter, remove the
flexible tube from the outlet of the model 90 and install the evacuated tedlar
bag. This assures that gaseous
components are undiluted by the air originally in the sampling system. Record requested data on the data form during
the sample time. The sampling is usually
done for 1 hour with a total of 5 to 6 liters sampled. When sampling is complete, record the precise
volume sampled. The process may require
different sample times or sample volumes.
(Sampling form is figure 25-8.)
4.1.5 Post
test leak check. Remove the tedlar bag
and replace it with the flexible tube.
Stopper the probe and operate the same as the pretest leak check
specified in section 4.1.3. If the leak
test is not acceptable, invalidate the sample.
4.2 Sample
recovery. The tedlar bag is ready for
direct analysis on the GC. The adsorber
tube shall undergo the following 2-stage preparation:
4.2.1 Sample
purge. The absorber tube is placed in
the Byron model 75 oven with a clean tedlar bag attached directly to the
tube. A volume of clean dry air is
passed through the adsorber tube while holding the oven temperature at about
130 degrees Celsius. The volume of air should
be precisely the same as that sampled.
This purge is necessary to remove any CO2 on the sample tube,
and the elevated temperature is needed to assure CO2 removal from
any absorbed water. The tedlar bag is
now ready for direct analysis on the GC.
4.2.2 Sample
digest. The
absorber tube, now free of CO2 and the lighter NMOs, is now attached
to an oxidation catalyst, and another tedlar bag is attached to the outlet of
the oxidation catalyst. A volume of
clean dry air equal to that sampled is passed through this system while the
temperature on the sample tube is brought up to 600 degrees Celsius. If the sampled volume was less than 3 liters,
a larger volume shall be used in the digestion to assure completion. Usually a multiple of precisely 1.5 or 2.0 of
the sampled volume is sufficient. This
third tedlar bag is now ready for direct analysis on the GC. If anything other than CO2 is
found in this bag, the model 75 oxidation catalyst is
probably in need of replacement. In this
case the test would be invalid and would have to be redone.
4.3 Analysis. Each of the 3 bags is analyzed on the
GC. Each bag should be analyzed as soon
as possible after being filled. At the
completion of analysis, the bags shall be cleaned by repeated fillings with
either clean air or nitrogen. Before
being used again, the bags shall be checked by filling with clean air and then
analyzed on the GC to assure zero concentrations of all analyzed
substances. All pertinent calibration,
performance, and operational checks in sections 4.4 and 5 of method 25 apply to
the Byron system.
6. Calculations.
6.1 Nomenclature.
C1 = Concentration
of sample bag, ppm C, (NMO converted to methane).
C2 = Concentration
of purge bag, ppm C, (NMO converted to methane).
C3 = Concentration
of digest bag, ppm C, (CO2 converted to methane).
C = Ppm C (NMO)
6.2-6.4 (Delete).
6.5 C1, C2, C3 calculated directly as:
![]()
6.6 C = C1 + C2 + C3
Delete figures 25.3, 25.4, 25.9, and 25.10 from
method 25. Amend figures 25.1 and 25.8
from method 25 to read as follows:
History: 1993 MR 4, Eff.

FIGURE 25.8. EXAMPLE FIELD DATA FORM
|
PLANT |
TIME
START |
TIME
STOP |
TEMP
AMB. |
TEMP
STACK |
BAG
# |
FILTER
# |
FLOW
RATE |
LEAK
CHECK |
||||
|
SOURCE |
Date |
Bag
PrePurge |
Liter
Sample |
Liter
Purge |
Liter
Digest |
Operator |
||||||
|
|
|
Integrated Values |
|
||||||||
|
|
THC |
NMHC |
CH4 |
CO |
CO2 |
THC |
NMHC |
CH4 |
CO |
CO2 |
cal# |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
SAMPLE |
|
|
|
|
|
|
|
|
|
|
|
|
BAG |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Range |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
PURGE |
|
|
|
|
|
|
|
|
|
|
|
|
BAG |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Range |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
DIGEST |
|
|
|
|
|
|
|
|
|
|
|
|
BAG |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Range |
|
|
|
|
|
|
|
|
|
|
|
R 336.2007 Alternate version of procedure L, referenced in R 336.2040(10).
Rule
1007. (1) The alternate version of procedure L is as follows:
1. Introduction.
1.1 Applicability. This procedure is applicable for determining
the input of volatile organic compounds (voc), measured as equivalent propane
as measured by a flame ionization instrument.
It is intended to be used as a segment in the development of liquid/gas
protocols for determining voc capture efficiency (ce)
for surface coating and printing operations.
1.2 Principle. The amount of voc introduced to the process
(l) is the sum of products of the weight (w) of each voc containing liquid
(ink, paint, solvent, or similar material) used and its voc content (v),
corrected for a response factor (rf) to allow the input to be calculated in
terms of propane, the same calibration gas used in the gaseous voc
measurements. A sample of each coating
used is distilled to separate the voc fraction. The distillate is used to prepare a known
standard for analysis by a flame ionization analyzer (fia), calibrated against
propane, to determine its rf.
2. Apparatus and reagents.
2.1 Liquid weight.
2.1.1 Balances/digital scales. To weigh drums of voc
containing liquids to within 0.2 lb.
2.1.2 Volume measurement apparatus
(alternative). Volume meters, flow
meters, density measurement equipment, or similar material, as needed to
achieve the same accuracy as direct weight measurements.
2.2 Response factor (rf)
determination (fia technique). The voc
distillation and tedlar gas bag generation systems apparatus are shown in
figures 1 and 2. The following equipment
is required:
2.2.1 Sample collection can. An appropriately sized metal can to collect
voc-containing materials. The can shall
be constructed in such a way that it can be grounded to the coating container.
2.2.2 Needle valves. To control gas flow.
2.2.3 Regulators. For fia, calibration, dilution, and sweep gas
cylinders.
2.2.4 Tubing and fittings. Teflon and stainless steel
tubing and fittings with diameters and lengths and sizes determined by
connection requirements of the equipment.
2.2.5 Thermometer. Capable of measuring the temperature of the
hot water and oil baths to within 1 degree Celsius.
2.2.6 Analytical balance. To measure plus or minus
0.01 mg.
2.2.7 Microliter syringe. 10-microliter size.
2.2.8 Vacuum and pressure manometers. 0 to 760 mm (0 to 30 in.)
hg. U-tube
manometer, vacuum or pressure.
2.2.9 Hot oil bath, with stirring hot
plate. Capable of
heating and maintaining a distillation vessel at 110 plus or minus 3 degrees
Celsius.
2.2.10 Vacuum/water aspirator. A device capable of drawing
a vacuum to within 20 mm hg from absolute.
2.2.11 Rotary evaporator system. Complete with folded inner coil, vertical
style condenser, rotary speed control, and teflon
sweep gas delivery tube with valved inlet.
Buchi rotavapor or equivalent.
2.2.12 Ethylene glycol cooling/circulating bath. Capable of maintaining the condenser coil
fluid at minus 10 degrees Celsius.
2.2.13 Dry gas meter. For the precise measurement
of dilution gas volume. It shall be calibrated to a primary standard,
either spirometer or bubble meter.
2.2.14 Activated charcoal/mole sieve trap. To remove any trace level of organics picked
up from the dry gas meter.
2.2.15 Gas coil heater. Sufficient length of
0.125-inch stainless steel tubing to allow heating of the dilution gas to near
the water bath temperature before entering the volatilization vessel.
2.2.16 Water bath, with stirring hot plate. Capable of heating and
maintaining a volatilization vessel and coil heater at a temperature of 100
plus or minus 5 degrees Celsius.
2.2.17 Volatilization vessel. 50-milliliter midget impinger fitted with a
septum top and loosely filled with glass wool to increase volatilization
surface.
2.2.18 Tedlar gas bag. Capable of holding 30 liters of gas, flushed
clean with zero air, leak tested and evacuated.
2.2.19 Cylinder of compressed zero air. Used to supply dilution air
for making the tedlar bag gas samples.
2.2.20 Cylinder of compressed thc free n2.
Used as sweep gas in the rotary evaporator system.
2.2.21 Organic concentration analyzer. An fia with a span
value of 1.5 times the expected concentration as propane; however, other span
values may be used if it can be demonstrated that they would provide more
accurate measurements. The fia
instrument shall be the same instrument used in the gaseous analyses adjusted
with the same fuel, combustion air, and sample backpressure (flowrate)
settings. The system shall be capable of
meeting or exceeding the following specifications:
2.2.21.1 Zero drift.
Less than plus or minus 3.0% of the span value.
2.2.21.2 Calibration drift. Less than plus or minus
3.0% of span value.
2.2.21.3 Calibration error. Less than plus or minus
5.0% of the calibration gas value.
2.2.22 Integrator/data acquisition system. An analog or digital device or computerized
data acquisition system used to integrate the fia response or compute the
average response and record measurement data.
The minimum data sampling frequency for computing average or integrated
values is 1 measurement value every 5 seconds.
The device shall be capable of recording average values at least once
per minute.
2.2.23 Chart recorder (optional). A chart recorder or similar device is
recommended to provide a continuous analog display of the measurement results
during the liquid sample analysis.
2.2.24 Calibration and other gases. For calibration, fuel, and combustion air, if
required, contained in compressed gas cylinders. All calibration gases shall be traceable to
NIST standards and shall be certified by the manufacturer to plus or minus 1%
of the tag value. Additionally, the
manufacturer of the cylinder should provide a recommended shelf life for each
calibration gas cylinder over which the concentration does not change more than
plus or minus 2% from the certified value.
For calibration gas values that are not generally available, alternative
methods for preparing calibration gas mixtures, such as dilution systems, may
be used with prior approval.
2.2.24.1 Fuel.
99.995% hydrogen, 40% hydrogen/60% helium, or 40%
hydrogen/ 60% nitrogen. The fia
manufacturer's recommended fuel shall be used.
An attempt shall be made to avoid fuels with oxygen to avoid an oxygen
synergism effect that reportedly occurs when oxygen concentration varies
significantly from a mean value.
2.2.24.2 Carrier gas. High purity air with less
than 1 ppm of organic material (as propane) or less than 0.1% of the span
value, whichever is greater.
2.2.24.3 Fia linearity calibration gases. Low-, mid-, and high-range
gas mixture standards with a nominal propane concentration of 20 to 30, 45 to
55, and 70 to 80% of the span value in air, respectively. Other calibration values and other span
values may be used if it can be shown that more accurate measurements would be
achieved.
2.2.24.4 System calibration gas. Gas mixture standard which contains propane
in air and which approximates the voc concentration expected for the tedlar gas
bag samples.
3. Determination of liquid input
weight. A capture efficiency test shall
consist of not less than 3 sampling runs.
Each run shall cover at least 1 complete production or processing cycle
or shall be at least 1 hour in duration. For automotive surface coating
operations, the sampling time per run shall be based on coating a minimum of 3
representative vehicles.
3.1 Weight difference. Determine the amount of material introduced
to the process as the weight difference of the feed material before and after
each sampling run. In determining the
total voc-containing liquid usage, account for all of the following:
(a) The initial (beginning) voc-containing liquid
mixture.
(b)
Any solvent added during the test run.
(c)
Any coating added during the test run.
(d)
Any residual voc-containing liquid mixture remaining at the end of the
sample run.
3.1.1 Identify all points where
voc-containing liquids are introduced to the process. To obtain an accurate measurement of
voc-containing liquids, start with an empty fountain, if applicable. After completing the run, drain the liquid in
the fountain back into the liquid drum, if possible, and weigh the drum
again. Weigh the voc-containing liquids
to plus or minus 0.5% of the total weight (full) or plus or minus 0.1% of the
total weight of voc-containing liquid used during the sample run, whichever is
less. If the residual liquid cannot be
returned to the drum, drain the fountain into a preweighed empty drum to
determine the final weight of the liquid.
3.1.2 If it is not possible to measure a
single representative mixture, then weigh the various
components separately, for example, if solvent is added during the sampling
run, weigh the solvent before it is added to the mixture. If a fresh drum of voc-containing liquid is
needed during the run, then weigh both the empty drum and the fresh drum.
3.2 Volume measurement
(alternative). If direct weight
measurements are not feasible, the tester may use volume meters, flow rate
meters, and density measurements to determine the weight of liquids that are
used if it can be demonstrated that the technique produces results equivalent
to the direct weight measurements. If a
single representative mixture cannot be measured, measure the components
separately.
4. Determination of voc content in
input liquids.
4.1 Collection of liquid samples.
4.1.1 Collect a 1-pint or larger sample of
the voc-containing liquid mixture at each application location at the beginning
and end of each test run. A separate
sample shall be taken of each VOC-containing liquid that is added to the
application mixture during the test run.
If a fresh drum is needed during the sampling run, then obtain a sample
from the fresh drum.
4.1.2 When collecting the sample, ground the
sample container to the coating drum.
Fill the sample container as close to the rim as possible to minimize
the amount of headspace.
4.1.3 After the sample is collected, seal
the container so the sample cannot leak out or evaporate.
4.1.4 Label the container to identify
clearly the contents.
4.2 Distillation of VOC.
4.2.1 Assemble the rotary evaporator as
shown in figure 1.
4.2.2 Leak check
the rotary evaporation system by aspirating a vacuum of approximately 20 mm hg
from absolute. Close up the system and
monitor the vacuum for approximately 1 minute.
If the vacuum falls more than 125 mm hg in 1 minute, repair leaks and repeat.
4.2.3 Deposit approximately 20 mls of the
sample (inks, paints, or similar material) into the rotary evaporation
distillation vessel.
4.2.4 Turn off the aspirator and gradually
apply a vacuum to the evaporator of within 20 mm hg.
4.2.5 Begin heating the vessel at a rate of
2 to 3 degrees Centigrade per minute, maintaining the vacuum specified in
4.2.3. Care shall be taken to prevent
material bumping from the distillation flask.
4.2.6 Continue heating until a temperature
of 110 degrees Centigrade is achieved and maintain this temperature for not
less than 10 minutes or until the sample has dried in the distillation flask.
4.2.7 Slowly introduce the n2
sweep gas through the purge tube and into the distillation flask, taking care
to maintain not less than 125 mm hg vacuum at all times.
4.2.8 Continue sweeping the remaining solvent
VOC from the distillation flask and condenser assembly for 10 minutes or until
all traces of condensed solvent are gone from the vessel and the still head.
4.2.9 Disassemble the apparatus and transfer
the distillate to a labeled sealed vial.
4.3 Preparation of VOC standard bag
sample.
4.3.1 Assemble the bag sample generation
system as shown in figure 2 and bring the water bath up to a near-boiling
temperature.
4.3.2 Inflate the tedlar bag and perform a
leak check on the bag.
4.3.3 Evacuate the bag and close the bag
inlet valve.
4.3.4 Record the current barometric
pressure.
4.3.5 Record the starting reading on the dry
gas meter, open the bag inlet valve, and start the dilution zero air flowing
into the tedlar bag at approximately 2 liters per minute.
4.3.6 The bag sample VOC concentration shall
be similar to the gaseous voc concentration measured in the exhaust gas
ducts. The amount of liquid voc required
can be approximated using the equations in section 6, the gaseous voc
measurement results in terms of propane, and an assumed response factor of
1.0. Let cc3 equal the
exhaust gas concentration in terms of propane and rf=1.0. Calculate cvoc. Let bv =
20 liters and calculate ml, the approximate quantity of liquid to be
used to prepare the bag gas sample.
4.3.7 Quickly withdraw an aliquot
(approximately 5 microliters) of sample from the distillate vial with the
microliter syringe and record its weight from the analytical balance to the
nearest 0.01 Mg.
4.3.8 Inject the contents of the syringe
through the septum of the volatilization vessel into the glass wool inside the
vessel.
4.3.9
Reweigh and record the tare weight of the now empty
syringe.
4.3.10 Record the pressure and temperature of
the dilution gas as it is passed through the dry gas meter, as shown in the
figure 2 diagram.
4.3.11 After approximately 20 liters of
dilution gas have passed into the tedlar bag, close the valve to the dilution
air source and record the exact final reading on the dry gas meter.
4.3.12 The gas bag is
then analyzed by fia within 1 hour of bag preparation in accordance with the
procedures contained in section 4.4.
4.4 Determination of VOC response
factor.
4.4.1 Start up the fia instrument using the
same settings as used for the gaseous VOC measurements.
4.4.2 Perform the fia analyzer calibration
and linearity checks according to the procedure in section 5.1. Record the responses to each of the
calibration gases and the back-pressure setting of the fia.
4.4.3 Connect the tedlar bag sample to the
fia sample inlet and record the bag concentration in terms of propane. Continue the analysis until a steady reading
is obtained for not less than 30 seconds.
Record the final reading and proceed with the calculation of the
response factor.
4.5 Determination of coating VOC content
as VOC (vu).
4.5.1 Determine the VOC content of the
coatings used in the process using EPA method 24 or 24a as applicable.
5. Calibration and quality
assurance.
5.1 Fia calibration and linearity
check. Make necessary adjustments to the
air and fuel supplies for the fia and ignite the burner. Allow the fia to warm up for the period
recommended by the manufacturer. Inject
a calibration gas into the measurement system and adjust the back-pressure
regulator to the value required to achieve the flow rates specified by the
manufacturer. Inject the zero- and the
high-range calibration gases and adjust the analyzer calibration to provide the
proper responses. Inject the low- and
mid-range gases and record the responses of the measurement system. The calibration and linearity of the system
are acceptable if the responses for all 4 gases are within 5% of the respective
gas values. If the performance of the
system is not acceptable, repair or adjust the system and repeat the linearity
check. Conduct a calibration and linearity
check after assembling the analysis system and after a major change is made to
the system. A calibration curve consisting of zero gas and 2 calibration levels
shall be performed at the beginning and end of each batch of samples.
5.2 Systems drift checks. After each sample, repeat the system
calibration checks in section 5.1 before any adjustments to the fia or
measurement system are made. If the zero
or calibration drift is more than plus or minus 3% of
the span value, discard the result and repeat the analysis.
5.3 Quality control. A minimum of 1 sample in each batch shall be
distilled and analyzed in duplicate as a precision control. If the results of the 2 analyzed differ by
more than plus or minus 10% of the mean, then the system shall be reevaluated
and the entire batch shall be redistilled and analyzed.
6. Calculations.
6.1 Bag sample volume, Bv.
|
|
Where:
Bv = Bag sample volume in standard liters.
Mv = Indicated dry gas meter volume, in
liters.
TSTD = 2930K.
TM = Meter gas temperature, in 0K.
PM = Meter gas pressure, in mm Hg absolute.
PSTD = 760 mm Hg.
6.2 Bag
sample voc concentration, as voc, Cvoc.
Cvoc
= Ml/Bv
Where:
Cvoc = Bag sample voc concentration, as voc,
mg/std. liters.
Ml = Weight of VOC liquid injected, mg.
6.3 Bag
sample VOC concentration, as propane, Cc3.
Cc3 = Rc3*K
Where:
Cc3 = Bag sample VOC concentration, as
propane, mg C3/std. liter.
Rc3 = FIA reading for bag gas sample, ppm
propane.
![]()
6.4 Response factor, RF.
RF = Cvoc/Cc3
Where:
RF = Response factor, weight VOC/weight
propane.
6.5 Total
voc content of the input voc containing liquid, as propane, L.
|
|
Where:
L
= Total VOC content of liquid input,
calculate as propane, kg.
VIJ
= Initial VOC weight fraction of VOC liquid
J.
VFJ
= Final VOC weight fraction of VOC liquid
J.
VAJ
= Voc weight fraction of voc liquid J
added during the test.
WIJ
= Weight of VOC containing liquid J at
beginning of test, kg.
WFG
= Weight of VOC containing liquid J at
end of test, kg.
WAJ
= Weight of VOC containing liquid J added
during the test, kg.
RFJ
= Response factor for VOC in liquid J,
weight VOC/weight propane.
History: 1993 MR 4, Eff.
FIGURE 1.
VOC Distillation Apparatus

Paint
Solvent Distillation Schematic
Response
Factor Determination Procedure
FIGURE 2.
Bag Sample Generation System

Sample
Preparation Equipment Schematic
Response Factor Determination Procedure
R 336.2010 Rescinded.
History: 1985 MR 2, Eff. Feb. 22, 1985; 1992 MR 9, Eff. Oct. 31, 1992; 1997 MR 5, Eff. May 26, 1997.
R 336.2011 Reference test method 5B.
Rule
1011. Reference test method 5B,
in-stack filtration method, reads as follows:
(a) The principle, applicability, and performance
test criteria are as follows:
(i) Principle.
Particulate matter is withdrawn isokinetically from the source and
collected on solid filtering media maintained at stack temperature. The particulate matter mass is determined
gravimetrically after removal of uncombined water.
(ii) Applicability. This method is applicable for the
determination of particulate emissions from stationary sources as identified in
table 31 of R 336.1331. The method
is also applicable when specifically provided for in the department’s rules,
orders, a permit to install, or a permit to operate.
(iii) Performance test criteria:
(A) A performance test shall consist of a minimum
of 3 separate samples of a specific air contaminant conducted within a 36-hour
period, unless otherwise authorized by the department. Each of the 3 separate samples shall be
obtained while the source is operating at a similar production level. For the
purpose of determining compliance with an applicable emission limit, rule, or
permit condition, the arithmetic mean of results of the 3 samples shall
apply. If a sample is accidentally lost
or conditions occur in which 1 of the 3 samples must be discontinued because of
forced shutdown, failure of an irreplaceable portion of the sampling train,
extreme meteorological conditions, or other circumstances beyond the owner's or
operator's control, compliance may, upon the approval of the department, be
determined using the arithmetic mean of the results of 2 samples.
(B) For any source that is subject to an emission
limitation calculated to 50% excess air, the multipoint, integrated sampling
procedure of R 336.2004(1)(c) shall be used for
gas analysis. For all other sources that require a
determination of the molecular weight of the exhaust, any optional sampling
procedure of R 336.2004(1)(c) may be used.
Alternatives or modifications to procedures are subject to the approval
of the department.
(C) The minimum volume per sample shall be 30
cubic feet of dry gas corrected to standard conditions (70 degrees Fahrenheit,
29.92 inches mercury). Minimum sample
time shall be 60 minutes, which may be continuous or a combination of shorter
sampling periods for sources that operate in a cyclic manner. Smaller sampling times or sample volumes, if
necessitated by process variables or other factors, may be approved by the
department.
(D) For any source whose emission control device
alters the moisture content of the exhaust gas, a moisture determination shall
be performed in a location upstream from the emission control device and in
accordance with R 336.2004(1)(d) or an
alternative method approved by the department.
(b) The following provisions apply to apparatus:
(i) Sampling train. A schematic of the sampling train used in
this method is shown in figure 102.
Construction details for many, but not all, of the train components are
given in APTD-0581. (See subdivision (g)(ii) of this rule.)
For changes from the APTD-0581 document and for allowable modifications
to figure 102, the user shall consult with the department.
The operating and
maintenance procedures for many, but not all, of the sampling train are
described in APTD-0576. (See subdivision
(g)(iii) of this rule.) Since correct usage is important in obtaining
valid results, all users shall read APTD-0576 and adopt the applicable
operating and maintenance procedures outlined in it, unless otherwise specified
herein. The sampling train shall consist
of the following components:
(A) Probe nozzle.
Stainless steel (316) or glass with sharp,
tapered leading edge. The angle of taper
shall be less than 30 degrees and the taper shall be on the outside to preserve
a constant internal diameter. The probe
nozzle shall be of the button-hook design, unless otherwise specified by the
department. If made of stainless steel,
the nozzle shall be constructed from seamless tubing. Other materials of construction may be used,
subject to the approval of the department.
A range of nozzle sizes
suitable for isokinetic sampling shall be available, for example, 0.32 to 1.27
cm (1/8 to 1/2 in. - or larger if higher volume sampling trains are used -
inside diameter (ID) nozzles in increments of 0.16 cm (1/16 in.). Each nozzle shall be calibrated according to
the procedures outlined in subdivision (e) of this rule.
(B) Probe liner.
Interior surface may be constructed of stainless steel (no specific
grade), glass, teflon, or such other material that
maintains proper flow at the stack conditions experienced.
(C) Pitot tube.
Type S, as described in section 2.1 of method 2, or other device
approved by the department. The pitot tube shall be attached to the probe, as shown in
figure 102, to allow constant monitoring of the stack gas velocity. The impact (high pressure) opening plane of
the pitot tube shall be even with or above the nozzle
entry plane (see method 2, figure 2-6b) during sampling. The type S pitot
tube assembly shall have a known coefficient, determined as outlined in section
4 of method 2.
(D) Differential pressure gauge. Incline manometer or equivalent devices (2)
as described in section 2.2 of method 2.
One manometer shall be used for velocity head ( p)
readings and the other shall be used for orifice differential pressure readings.
(E) Filter holders. Two separate filter holders in series or 1
filter holder with separate filter supports and seals for 2 filters. One filter holder with 2 filters held in
contact with each other is not acceptable.
Materials of construction may be stainless steel (316), glass, teflon, or other material approved by the department.
(F) Filter heating system. Auxiliary heating of the filter media is not
acceptable. For saturated stack gases,
the operator may opt to use filters that do not blind when wet and that do not
require heating (see subdivision (c)(i)(A) of this
rule).
(G) Condenser.
The following system shall be used to determine the stack gas moisture
content: Three impingers connected in
series with leak-free ground glass fittings or any similar leak-free
noncontaminating fittings. All impingers
shall be of the Greenburg-Smith design and shall be modified by replacing the
tip with a 1.3 cm (1/2 in.) ID glass tube extending to about
1.3 cm (1/2 in.) from the bottom of the flask. Modifications, such as using flexible
connections between the impingers or using materials other than glass, are
permitted, subject to the approval of the department. The first impinger shall contain a known
quantity of water (subdivision (d)(i)(C) of this rule);
the second shall be empty; and the third shall contain a known weight of silica
gel or equivalent desiccant.
Alternatively, any
system that cools the sample gas stream and
allows measurement of the water condensed and moisture leaving the
condenser, each to within 1 ml or 1 g, may be used, subject to the approval of
the department. In any case, the means
for measuring the moisture leaving the condenser shall be by passing the sample
gas stream through a tared silica gel, or equivalent desiccant, trap with exit
gases kept below 20 degrees Centigrade (68 degrees Fahrenheit) and determining
the weight gain.
If a determination of
the particulate matter collected in the impingers is required by the
department’s rules, a permit to install, or a permit to operate, then the
impinger system described above shall be used without modification. Contact the department as to the sample
recovery and analysis of the impinger contents.
(H) Metering system. Vacuum gauge, leak-free pump, thermometers
capable of measuring temperature to within 3 degrees Centigrade (5.4 degrees Fahrenheit), dry-gas meter
capable of measuring volume to within 2%, and related equipment as shown in
figure 102. Other metering systems
capable of maintaining sampling rates within 10% of isokinetic and capable of
determining sample volumes to within 2% may be used, subject to the approval of
the department. When the metering system
is used in conjunction with a pitot tube, the system shall enable checks of
isokinetic rates.
Sampling trains
utilizing metering systems designed for higher flow rates than those described
in APTD-0581 or APTD-0576 may be used if the specifications of this method are
met.
(I) Barometer. Mercury, aneroid, or other
barometer capable of measuring atmospheric pressure to within 2.5 mm Hg (0.1
in. Hg).
In many cases, the barometric reading may be obtained from a nearby
national weather service station. In
this case, the station value, which is the absolute barometric pressure, shall
be requested and an adjustment for elevation differences between the weather
station and sampling point shall be applied at a rate of minus 2.5 mm Hg (0.1
in. Hg) per 30 M (100 ft.) elevation increase or vice versa
for elevation decrease.
(J) Gas density determination equipment. Temperature sensor and
pressure gauge, as described in sections 2.3 and 2.4 of method 2, and gas
analyzer, if necessary, as described in method 3. The temperature sensor shall, preferably, be
permanently attached to the pitot tube or sampling
probe in a fixed configuration such that the tip of the sensor extends beyond
the leading edge of the probe sheath and does not touch any metal. Alternatively, the sensor may be attached
just before use in the field. If the
temperature sensor is attached in the field, then the sensor shall be placed in
an interference-free arrangement with respect to the type S pitot
tube openings (see method 2, figure 2-7).
As a second alternative, if a difference of not more than 1% in the
average velocity measurement is to be introduced, then the temperature gauge
need not be attached to the probe or pitot tube. This alternative is subject to the approval
of the department.
“Construction Details
of Isokinetic Source Sampling Equipment,” APTD-0581,
April 1971 (PB203-060-LL), and
“Maintenance, Calibration, and
Operation of Isokinetic Source Sampling Equipment,” APTD-0576, March 1972
(PB209-022-LL), are adopted by reference in this rule. Copies of these documents may be inspected at
the
(ii) Sample recovery. The following items are required:
(A) Probe-liner and probe-nozzle brushes. Nylon bristle brushes with stainless steel
wire handles. The probe brush shall have
extensions, at least as long as the probe, made of
stainless steel, nylon, teflon, or similarly inert material. The brushes shall be properly sized and
shaped to brush out the probe liner and nozzle.
(B) Wash bottles -- 2. Glass wash bottles are recommended; the
tester may use polyethylene wash bottles, but the acetone should not be stored
in polyethylene bottles for longer than 1 month.
(C) Glass sample storage containers. Chemically resistant,
borosilicate glass bottles for acetone washes, 500 ml or 1000 ml. Screw cap liners shall either be
rubber-backed teflon or shall be constructed so as to
be leak-free and resistant to chemical attack by acetone. Narrow-mouth glass bottles are less prone to
leakage. Alternatively, polyethylene
bottles may be used.
(D) Filter containers. Glass, polyethylene, or
aluminum tube containers, unless otherwise specified by the department.
(E) Graduated cylinder or balance. To measure condensed water to within 1 ml or
1 g., graduated cylinders shall have subdivisions of not more than 2 ml. Most laboratory balances are capable of
weighing to the nearest 0.5 g or less.
Any of these balances may be used here and in paragraph (iii)(D) of this subdivision.
(F) Plastic storage containers. Airtight containers to store silica gel.
(G) Funnel and rubber policeman. To aid in the transfer of
silica gel to container; not necessary if silica gel is weighed in the field.
(H) Funnel.
Glass or polyethylene, to aid in sample recovery.
(iii) Analysis.
The following equipment is required for analysis:
(A) Glass weighing dishes.
(B) Desiccator.
(C) Analytical balance. To measure to within 0.1
mg.
(D) Balance.
To measure to within 0.5 mg.
(E) Beakers.
250 ml.
(F) Hygrometer.
To measure the relative humidity of the laboratory
environment.
(G) Temperature gauge. To measure the temperature
of the laboratory environment.
(c) The following provisions shall apply to
reagents:
(i) Sampling.
The reagents used in sampling are as follows:
(A) Filters. Two in-stack filters may be any
combination of alundum ceramic thimble filters, type RA-98, or glass fiber
filters, type A without organic binder.
The size of such filters shall allow proper sampling rates to maintain
isokinetics using the nozzle sizes specified in subdivision (b)(i)(A) of this rule.
Alternatively, other
types of filters may be used, subject to the approval of the department.
(B) Silica gel.
Indicating type, 6 to 16 mesh. If
previously used, dry at 175 degrees Centigrade (350 degrees Fahrenheit) for 2
hours. New silica gel may be used as
received. Alternatively, other types of
desiccants that are equivalent or better may be used, subject to the approval
of the department.
(C) Water.
When analysis of the material caught in the impingers is required,
distilled water shall be used. Run
blanks before field use to eliminate a high blank of test samples.
(D) Crushed ice.
(E) Stopcock grease. Acetone-insoluble,
heat-stable silicone grease. This
is not necessary if screw-on connectors with teflon
sleeves, or equivalent, are used.
Alternatively, other types of stopcock grease may be used, subject to
the approval of the department.
(ii) Sample recovery. Washing solvent. Either acetone or distilled water may be used
for sample recovery. If acetone is used
for washing solvent, then reagent grade, less than 0.001% residue, in glass
bottles is required. Acetone from metal
containers generally has a high residue blank and shall not be used.
If suppliers transfer
acetone to glass bottles from metal containers, then acetone blanks shall be
run before field use and only acetone with low blank values (less than 0.001%)
shall be used. In no case shall a blank
value of more than 0.001% of the weight of acetone used be subtracted from the
sample weight.
If distilled water is
used for washing solvent, use distilled water with less than 0.001%
residue. Run blanks before field use to
eliminate a high blank on test samples.
(iii) Analysis.
Two reagents are required for the analysis:
(A) Solvent.
Same as paragraph (ii) of this subdivision for
quantitative transfer.
(B) Desiccant. Anhydrous
calcium sulfate, indicating type. Alternatively, other types of
desiccants may be used, subject to the approval of the department.
(d) The following provisions shall apply to
procedure:
(i) Sampling.
The complexity of this method is such that, in order to obtain reliable
results, testers shall be trained and experienced with the test
procedures. Sampling shall comply with
the following provisions:
(A) Pretest preparation. All the components shall be maintained and
calibrated according to the applicable procedures described in APTD-0576, unless otherwise specified in this rule.
Weigh several 200 to
300 g portions of silica gel in airtight containers to the nearest 0.5 g. Record the total weight of the silica gel
plus container on each container. As an
alternative, the silica gel need not be preweighed, but may be weighed directly
in its impinger or sampling holder just before train assembly.
Check filters visually
against light for irregularities, flaws, pinhole leaks, or cracks. Label filters of the proper size on the back
side using numbering machine ink. As an
alternative, label the shipping containers (subdivision (b)(ii)(D)
of this rule) and keep the filters in these containers at all times, except
during sampling and weighing.
Dry the filters in an
oven at 105 degrees Centigrade (220 degrees Fahrenheit) for a minimum of 2
hours, cool for at least 1 hour in a desiccator containing anhydrous calcium
sulfate, and individually weigh and record their weights to the nearest 0.1
mg. During the weighing, the filter
shall not be exposed to the laboratory atmosphere for a period of more than 2
minutes and a relative humidity above 50%.
Procedures, other than those specified, that account for relative
humidity effects may be used, subject to the approval of the department.
(B) Preliminary determinations. Select the sampling site and the minimum
number of sampling points according to method 1 or as specified by the
department. Determine the stack
pressure, temperature, and the range of velocity heads using method 2; it is
recommended that a leak check of the pitot lines (see method 2, section 3.1) be
performed. Determine the moisture
content using approximation method 4, or its alternatives, for the purpose of
making isokinetic sampling rate settings.
Determine the stack gas dry molecular weight, as described in method 2,
section 3.6; if integrated method 3 sampling is used for molecular weight
determination, the integrated bag sample shall be taken simultaneously with,
and for the same total length of time as, the particulate sample run.
Select a nozzle size
based on the range of velocity heads so that it is not necessary to change the
nozzle size to maintain isokinetic sampling rates. During the run, do not change the nozzle
size. Ensure that the proper
differential pressure gauge is chosen for the range of velocity heads
encountered (see section 2.2 of method 2).
Select a suitable
probe liner and probe length so that all traverse points may be sampled. For large stacks, sampling from opposite
sides of the stack may reduce the length of probes.
Select a total sampling
time greater than or equal to the minimum total sampling time specified in the
department’s rules so that the sampling time per point is not less than 5
minutes, unless approved by the department, or some greater time interval as
specified by the department, and so that the sample volume taken, corrected to
standard conditions, exceeds the required minimum total gas sample volume. The latter is based on an approximate average
sampling rate.
The number of minutes
sampled at each point may be an integer or an integer plus 1/2 minute to avoid
timekeeping errors.
In some circumstances,
such as in batch cycles, it may be necessary to sample for shorter times at the
traverse points and to obtain smaller gas sample volumes. In these cases, the department’s approval
shall first be obtained.
(C) Preparation of collection train. During preparation and assembly of the
sampling train, keep all openings where contamination can occur covered until
just before assembly or until sampling is about to begin.
Place 100 ml of water
in the first impinger, leave the second impinger empty, and transfer
approximately 200 to 300 g of preweighed silica gel from its container to the
third impinger. More silica gel may be
used, but care shall be taken to ensure that it is not entrained and carried
out from the impinger during sampling.
Place the container in a clean place for later use in the sample
recovery. Alternatively, the weight of
the silica gel plus impinger may be determined to the nearest 0.5 g and
recorded.
Using tweezers or
clean disposable surgical gloves, place a labeled (identified) and weighed
filter in each filter holder. Be sure
that the filter is properly centered and the gasket properly placed so as to
prevent the sample gas stream from circumventing the filter.
Install the selected
nozzle using a Viton A 0-ring when stack temperatures are less than 260 degrees
Centigrade (500 degrees Fahrenheit) and an asbestos string gasket when
temperatures are higher. See APTD-0576 for requirements. Other connecting systems using either 310
stainless steel or teflon ferrules may be used to form
a leak-free direct mechanical connection.
Mark the probe with heat-resistant tape or by some other method to
denote the proper distance into the stack or duct for each sampling point.
Set up the train as in
figure 102. If necessary, use a very
light coat of silicone grease on all ground glass joints. Grease only the outer portion (see APTD-0576)
to avoid the possibility of contamination by the silicone grease.
Place crushed ice
around the impingers.
(D) Leak check
procedures:
(1) Pretest leak check. A pretest leak check
is strongly recommended, but not required, to prevent invalid sampling and
wasted time. If the tester opts to
conduct the pretest leak check, the following procedure shall be used:
Perform the leak check
on the entire system, including filter housings and nozzle, by plugging the
nozzle and pulling a 380 mm Hg (15 in. Hg) vacuum.
Alternatively, a lower vacuum may be used if it is not exceeded during the test. Leakage rates in excess of 4% of the average
sampling rate or 0.00057 m³/min (0.02 cfm), whichever is
less, are unacceptable.
The following leak
check instructions for the sampling train described in APTD-0576 and APTD-0581
may be helpful. Start the pump with the
bypass valve fully open and the coarse adjust valve completely closed. Partially open the coarse
adjust valve and slowly close the bypass valve until the desired vacuum is
reached. Do not reverse the direction of
the bypass valve; this will cause water to back up into the flexible sample
tube and the probe. If the desired
vacuum is exceeded, either leak-check at this higher vacuum or end the leak check and start over.
When the leak check is
completed, first slowly remove the plug from the inlet to the nozzle and
immediately turn off the vacuum pump.
This prevents the water in the first impinger from being forced backward
into the sample tube and prevents silica gel from being entrained backward into
the second impinger.
(2) Leak checks during sample run. If, during the sampling run, a component
(such as a filter assembly or impinger) change becomes necessary, a leak check
shall be conducted immediately before the change is made. The leak check shall
be done according to the procedure outlined in paragraph (i)(D)(1) of this
subdivision, except that it shall be done at a vacuum equal to or greater than
the maximum value recorded up to that point in the test. If the leakage rate is not more than 0.00057
m³/min (0.02 cfm) or 4% of the average sampling rate, whichever is less, then
the results are acceptable and no correction need be applied to the total
volume of dry gas metered. If a higher
leakage rate is obtained, then the tester shall either record the leakage rate
and plan to correct the sample volume, as shown in subdivision (f)(iii) of this rule, or shall void the sampling run.
Immediately after
component changes, leak checks may be performed. If leak checks are done, then the procedure
outlined in paragraph (i)(D)(1) of this subdivision
shall be used.
(3) Post-test leak
check. A leak check is required at the
conclusion of each sampling run. The
leak check shall be performed in accordance with the procedures in paragraph (i)(D)(1) of this subdivision, except that it shall be conducted
at a vacuum equal to or greater than the maximum value reached during the
sampling run. If the leakage rate is not
more than 0.00057 m³/min (0.02 cfm) or 4% of the average sampling rate,
whichever is less, then the results are acceptable and no correction need be
applied to the total volume of dry gas metered.
If a higher leakage rate is obtained, then the tester shall either
record the leakage rate and correct the sample volume, as shown in subdivision
(f)(iii) of this rule, or shall void the sampling run.
(E) Particulate train operation. During the sampling run, maintain an
isokinetic sampling rate that is within 10% of true isokinetic, unless otherwise
specified by the department.
For each run, record
the data required on a data sheet such as the data sheet in figure 104. Record the initial dry-gas meter
reading. Record the dry-gas meter
readings at the beginning and end of each sampling time increment, when changes
in flow rates are made, before and after each leak check, and when sampling is
halted. Take other readings required by
figure 104 at least once at each sample point during each time increment, and
take additional readings when significant changes (20% variation in velocity
head readings) necessitate additional adjustments in flow rate. Level and zero the manometer. Because the manometer level and zero may
drift due to vibrations and temperature changes, make periodic checks during
the traverse.
Clean the portholes
before the test run to minimize the chance of sampling deposited material. To begin sampling, remove the nozzle cap and
verify that the pitot tube and probe are properly
positioned. Position the nozzle at the
first traverse point with the tip pointing directly into the gas stream. Immediately start the pump and adjust the
flow to isokinetic conditions.
Nomographs that aid in the rapid adjustment of the isokinetic sampling
rate without excessive computations are available. These nomographs are designed for use when
the type S pitot tube coefficient is 0.85 ±0.02 and
the stack gas equivalent density (dry molecular weight) is equal to 29 ±4. APTD-0576 details the procedure for using the
nomographs. If Cp and Md
are outside the above stated ranges, do not use the nomographs unless
appropriate steps (subdivision (g)(iv) of this rule)
are taken to compensate for the deviations.
When the stack is
under significant negative pressure (height of impinger stem), take care to
pull low flow when inserting the probe into the stack to prevent water from
backing into the sample tubing and to avoid pulsation through the filter and
possible loss of materials.
When the probe is in
position, block off the openings around the probe and porthole to prevent
unrepresentative dilution of the gas stream.
Traverse the stack
cross section, as required by method 1 or as specified by the department, being
careful not to bump the probe nozzle into the stack walls when sampling near
the walls or when removing or inserting the probe through the portholes; this
minimizes the chance of extracting deposited material.
During the test run,
add more ice and, if necessary, salt to maintain a temperature of less than 20
degrees Centigrade (68 degrees Fahrenheit) at the condenser/silica gel
outlet. Also, periodically check the
level and zero of the manometer.
If the pressure drop
across the filter becomes too high and makes isokinetic sampling difficult to
maintain, the filter may be replaced in the midst of a sample run. It is recommended that another complete
filter assembly be used rather than attempting to change the filter
itself. Before a new filter assembly is
installed, conduct a leak check (paragraph (i)(D)(2)
of this subdivision). The total
particulate weight shall include the summation of all filter assembly catches.
A single train shall be
used for the entire sample run, except in cases where simultaneous sampling is
required in 2 or more separate ducts, at 2 or more different locations within
the same duct, or where equipment failure necessitates a change of trains. In all other situations, the use of 2 or more
trains shall be subject to the approval of the department.
When 2 or more trains
are used, separate analyses of the front-half and, if applicable, impinger
catches from each train shall be performed, unless identical nozzle sizes were
used on all trains. If identical nozzle
sizes were used, the front-half catches from the individual trains may be
combined, as may the impinger catches, and 1 analysis of front-half catch and 1
analysis of impinger catch may be performed.
Consult with the department for details concerning the calculation of
results when 2 or more trains are used.
At the end of the
sample run, turn off the coarse adjust valve, remove the probe and nozzle from
the stack, turn off the pump, record the final dry-gas meter reading, and
conduct a post-test leak check, as outlined in paragraph (i)(D)(3) of this
subdivision. Leak-check the pitot lines
as described in method 2, section 3.1; the lines shall pass this leak check to
validate the velocity head data.
(F) Calculation of percent isokinetic. Calculate percent isokinetic (see subdivision
(f) of this rule) to determine if the run was valid or if another test run
should be made. If there was difficulty
in maintaining isokinetic rates due to source conditions, consult with the
department for possible variance on the isokinetic rates.
(ii) Sample recovery. Proper cleanup procedure begins as soon as
the probe is removed from the stack at the end of the sampling period. Allow the probe to cool.
When the probe can be
safely handled, wipe off all external particulate matter near the tip of the
probe nozzle and place a cap over it to prevent losing or gaining particulate
matter. Do not cap off the probe tip
tightly while the sampling train is cooling down as this creates a vacuum in
the filter holder and draws water from the impingers into the sample tube.
Before moving the
sampling train to the cleanup site, make sure all condensed water in the probe
and flexible sample lines are drained into the first impinger. Disconnect all sample lines and remove the
nozzle-filter set assembly from the probe.
Cap all openings to prevent contamination or accidental loss of
sample. Remove all excess particulate
from the exterior of the nozzle-filter assembly to prevent contamination during
disassembly.
Transfer the
nozzle-filter set assembly and impinger set to the cleanup area. The cleanup area shall be clean and protected
from the wind so that the chances of contaminating or losing the sample are
minimized.
Save a portion of the
solvent used for cleanup as a blank.
Take 200 ml of this solvent directly from the wash bottle being used and
place it in a glass sample container labeled "solvent blank".
Inspect the train
before and during disassembly and note any abnormal conditions. Treat the samples in the following manner:
Container
Nos. 1, 1A. Carefully remove the
filters from the filter holders and place in their identified containers. Use a pair of tweezers or clean disposable
surgical gloves, or both, to handle the filters. Carefully transfer to the container any
particulate matter or filter fibers, or both, that adhere to the filter holder
gasket by using a dry nylon bristle brush or sharp-edged blade, or both. Seal the containers.
Container
No. 2. Taking care to see that
particulate on the outside of the nozzle and filter holders does not get into
the sample, the tester shall carefully remove the nozzle and clean the inside
surface by rinsing with solvent from a wash bottle and brushing with a nylon
bristle brush. Brush until the solvent rinse shows no visible particles and
then make a final rinse of the inside surface with solvent.
After ensuring that
all joints have been cleaned of all extraneous material, the tester shall
quantitatively remove particulate from the filter holders by rubbing the
surfaces with a nylon bristle brush and rinsing with solvent. Rinse each surface 3 times, or more if needed, to remove visible particulate. Make a final rinse of the brush and filter
holder set. After all solvent washings and
particulate matter have been collected in the sample container,
tighten the lid on the sample container so that solvent will not leak out when
it is shipped to the laboratory. Mark
the height of the fluid level to determine if leakage occurred during transport. Label the container to clearly identify its
contents.
Container
No. 3. Note the color of the
indicating silica gel to determine if it has been completely spent and make a
notation of its condition. Transfer the
silica gel from the third impinger to its original container and seal. A funnel may make it easier to pour the
silica gel without spilling it. A rubber
policeman may be used as an aid in removing the silica gel from the
impinger. It is not necessary to remove
the small amount of dust particles that adhere to the impinger wall and are
difficult to remove. Since the gain in
weight will be used for moisture calculations, do not use any water or other
liquids to transfer the silica gel. If a
balance is available in the field, follow the procedure for container No. 3 in
paragraph (iii) of this subdivision.
Impinger
water. Treat the impingers in the
following manner: Make a notation of any
color or film in the liquid catch.
Measure the liquid that is in the first 2 impingers to within ±1 ml by
using a graduated cylinder or by weighing it to within ±1.0 g by using a
balance if one is available. Record the
volume or weight of liquid present. This
information is required to calculate the moisture content of the effluent gas.
Discard the liquid
after measuring and recording the volume or weight, unless analysis of the
impinger catch is required (subdivision (b)(i)(G) of
this rule).
If a different type
of condenser is used, measure the amount of moisture condensed either
volumetrically or gravimetrically.
If possible,
containers shall be shipped in a manner that keeps them upright at all times.
(iii) Analysis.
Record the data required on a sheet such as the sheet in figure
106. Handle each sample container in the
following manner:
Container
Nos. 1, 1A. Analyze and report
each filter separately. Transfer the
filter and any loose particulate from the sample container to a tared-glass
weighing dish. Dry the filter in an oven
at 105 degrees Centigrade (220 degrees Fahrenheit) for a minimum of 2 hours,
cool for at least 1 hour in a desiccator containing anhydrous calcium sulfate
and weigh and record its weight to the nearest 0.1 mg. During the weighing the filter shall not be
exposed to the laboratory atmosphere for a period greater than 2 minutes or a
relative humidity above 50%. Procedures,
other than those specified, that account for relative humidity effects may be
used, subject to the approval of the department. The method used for drying and weighing of
filters shall be consistent before and after the test.
Container
No. 2. Note the level of liquid
in the container and confirm on the analysis sheet if leakage occurred during
transport. If a noticeable amount of
leakage has occurred, then either void the sample or
use methods, subject to the approval of the department, to correct the final
results. Measure the liquid in this
container either volumetrically to ±1 ml or gravimetrically to ±1.0 g. Transfer the contents to a tared 250 ml
beaker and evaporate to dryness either at ambient temperature and pressure for
acetone or at 95 degrees Centigrade (203 degrees Fahrenheit) in an oven for
distilled water. Then subject the sample
to 250 degrees Centigrade (482 degrees Fahrenheit) in an oven for 2 to 3
hours. Desiccate 24 hours and weigh to a
constant weight. Report the results to
the nearest 0.1 mg.
Container
No. 3. Weigh the spent silica
gel, or silica gel plus impinger, to the nearest 0.5 g using a balance. This step may be conducted in the field.
"Solvent
blank" container. Measure
solvent in this container either volumetrically or gravimetrically. Transfer the contents to a tared 250 ml
beaker and evaporate to dryness either at ambient temperature and pressure for
acetone or at 95 degrees Centigrade (203 degrees Fahrenheit) in an oven for
distilled water. Then subject the sample
to 250 degrees Centigrade (482 degrees Fahrenheit) in an oven for 2 to 3
hours. Desiccate for 24 hours and weigh
to a constant weight. Report the results
to the nearest 0.1 mg.
If acetone is used,
the contents of Container No. 2, as well as the acetone blank container, may be
evaporated at temperatures higher than ambient.
If evaporation is done at an elevated temperature, then the temperature
shall be closely supervised, and the contents of the beaker shall be swirled
occasionally to maintain an even temperature.
Use extreme care, as acetone is highly flammable and has a low flash
point.
(e) Calibration.
Maintain a laboratory log of all calibrations. Calibrations shall comply
with the following provisions:
(i) Probe nozzle.
Probe nozzles shall be calibrated before their initial use in the
field. Using a micrometer, measure the
inside diameter of the nozzle to the nearest 0.025 mm (0.001 in.). Make 3 separate measurements using different
diameters each time and obtain the average of the measurements. The difference between the high and low
numbers shall not exceed 0.1 mm (0.004 in.).
When nozzles become nicked, dented, or corroded, they shall be reshaped,
sharpened, and recalibrated before use.
Each nozzle shall be permanently and uniquely identified.
(ii) Pitot tube.
The type S pitot tube assembly shall be
calibrated according to the procedures in section 4 of method 2.
(iii) Metering system. Before its initial use in the field, the
metering system shall be calibrated according to the procedure in
APTD-0576. Instead of physically
adjusting the dry-gas meter dial readings to correspond to the wet-test meter
readings, calibration factors may be used to mathematically correct the gas
meter dial readings to the proper values.
Before calibrating the metering system, a leak check may be
conducted. For metering systems having
diaphragm or rotary pumps, the normal leak check procedure will not detect
leakages within the pump. For these
cases, the following leak check procedure may be used: Make a 10-minute calibration run at 0.00057
m³/min (0.02 cfm); at the end of the run, take the difference of the measured
wet-test meter and dry-gas meter volumes; divide the difference by 10 to get
the leak rate. The leak rate shall not
exceed 0.00057 m³/min (0.02 cfm).
After each field use,
the calibration of the metering system shall be checked by performing 3
calibration runs at a single, intermediate orifice setting, based on the
previous field test, with the vacuum set at the maximum value reached during
the test series. To adjust the vacuum,
insert a valve between the wet-test meter and the inlet of the metering system. Calculate the average value of the
calibration factor. If the calibration
has changed by more than 5%, then recalibrate the meter over the full range of
orifice settings, as outlined in APTD‑0576.
Alternatively, a
spirometer may be substituted for a wet-test meter in the above calibration
procedures.
Alternative procedures,
such as using the orifice meter coefficients, may be used, subject to the
approval of the department. If the
dry-gas meter coefficient values obtained before and after a test series differ
by more than 5%, then the test series shall be performed using whichever meter
coefficient value (before or after) gives the lower value of total sample
volume.
(iv) Temperature gauges. Use the procedure in section 4.3 of method 2
to calibrate in-stack temperature gauges.
Dial thermometers, such as those used for the dry-gas meter and
condenser outlet, shall be calibrated against mercury-in-glass thermometers.
(v) Leak check of metering system shown in figure
102. That portion of the sampling train from the pump to the orifice meter
shall be leak-checked before initial use and after each shipment. Leakage after the pump will result in less
volume being recorded than is actually sampled.
The following procedure is suggested (see figure 107): Close the main valve on the meter box. Insert
a 1-hole rubber stopper with rubber tubing attached into the orifice exhaust
pipe. Disconnect and vent the low side
of the orifice manometer. Close off the
low side orifice tap. Pressurize the
system to 13 to 18 cm (5 to 7 in.) water column by blowing into the rubber
tubing. Pinch off the tubing and observe
the manometer for 1 minute. A loss of
pressure on the manometer indicates a leak in the meter box; leaks, if present,
shall be corrected.
(vi) Barometer. Calibrate against a mercury barometer.
(f) Calculations.
When carrying out calculations, retain at least 1 extra decimal figure
beyond that of the acquired data. Round
off figures after the final calculation.
Other forms of the equations may be used if they give equivalent
results. The following provisions apply
to calculations:
(i) Nomenclature:
An = Cross-sectional
area of nozzle, m²(ft.²).
A = Cross-sectional
area of stack or flue at the point of sampling, ft².
Bws = Water
vapor in the gas stream, proportion by volume, expressed as a fraction.
Bwi = Percent
water vapor in gas entering source particulate control device determined by
method 4.
Bwo = Percent
water vapor in gas exiting source particulate control device.
Ca =
Cs = Concentration
of particulate matter in stack gas, pounds per 1,000 pounds of actual stack
gas.
CsD = Concentration
of particulate matter in stack gas, moisture excluded, pounds per 1000 pounds
of dry stack gas.
Cs50 = Concentration
of particulate matter corrected to 50% excess air, pounds per 1000 pounds of
stack gas.
Cs50D = Concentration
of particulate matter corrected to 50% excess air, excluding any water addition
from a collector, pounds per 1000 pounds of stack gas.
E = Mass
emission rate of particulate, lb/hr.
F50 = Concentration
conversion factor to 50% excess air with no moisture alterations in exhaust.
F50D = Concentration
conversion factor to 50% excess air, excluding any moisture added to exhaust
gas by pollution collection system.
FD = Concentration
conversion factor to dry basis, excluding any water in the stack gas.
I = Percent of isokinetic sampling.
La = Maximum
acceptable leakage rate for either a pretest leak check or for a leak check
following a component change; equal to 0.00057 m³/min (0.02 cfm) or 4% of the
average sampling rate, whichever is less.
Li = Individual
leakage rate observed during the leak check conducted before the
"ith" component change (i = 1, 2, 3 . . . . n), m³/min (cfm).
Lp = Leakage rate observed during the post-test leak
check, m³/min (cfm).
Md = Molecular
weight of dry stack gas, g/g mole (lb/lb-mole), calculated by method 3,
equation 3-2, using data from integrated method 3.
mn = Total amount of particulate matter collected,
mg.
Mw = Molecular
weight of water, 18.0 g/g-mole (18.0 lb/lb-mole).
ma = Mass of residue of solvent after evaporation,
mg.
mg = Total weight of gas samples through nozzle, lb.
Pbar = Barometric
pressure at the sampling site, mm Hg (in. Hg).
Ps = Absolute
stack gas pressure.
Pstd = Standard
absolute pressure, 760 mm Hg (29.92 in. Hg).
R = Ideal
gas constant, 0.06236 mm Hg-m³/°K-g-mole (21.85 in. Hg-ft.³/°R‑lb-mole).
Tm = Absolute
average dry-gas meter temperature (see figure 104), °K (°R).
Ts = Absolute
average stack gas temperature (see figure 104), °K (°R).
Tstd = Standard
absolute temperature, 294.I°K (530°R).
Va = Volume of solvent blank, ml.
Vaw = Volume
of solvent used in wash, ml.
Vlc = Total
volume of liquid collected in impingers and silica gel (see figure 106), ml.
Vm = Volume
of gas sample as measured by the dry-gas meter, dcm
(dcf).
Vm(std) = Volume of gas sample measured by the dry-gas
meter, corrected to standard conditions, dscm (dscf).
Vw(std) = Volume of water vapor in the gas sample,
corrected to standard conditions, scm (scf).
Vs = Stack
gas velocity, calculated by method 2, equation 2-9, using data obtained from
method 5, m/sec (ft./sec).
Wa = Weight of residue in solvent wash, mg.
Y = Dry-gas
meter calibration factor.
DH = Average pressure differential across the
orifice meter (see figure 104), mm H20 (in. H20).
%02 = Percent
oxygen in stack gas by volume (dry basis).
%N2 = Percent
nitrogen in stack gas by volume (dry basis).
pa = Density of solvent, mg/ml.
ps(std) = Density of all sampled gas at standard
conditions, lb/ft.³
pw = Density of water, 0.9982 g/ml (0.002201 lb/ml).
q = Total sample time, min.
q1 = Sample time, interval, from the beginning of a
run until the first component change, min.
qi = Sampling time interval, between 2 successive
component changes, beginning with the interval between the first and second
changes, min.
qp = Sampling time interval, from the final (nth)
component change until the end of the sampling run, min.
13.6 = Specific
gravity of mercury.
60 = Sec/min.
100 = Conversion
to percent.
386.9 = Cubic
feet per lb-mole of ideal gas at standard conditions.
453.6 = Conversion
of pounds to grams.
3600 = Conversion
of hours to sec.
1000 = Conversion
of 1000 lb units to lb units.
(ii) Average the dry-gas meter temperature and
average the orifice pressure drop. See
data sheet (figure 5-2).
(iii) Dry gas volume. Correct the sample volume measured by the
dry-gas meter to standard conditions (21.11 degrees Centigrade, 760 mm Hg or 70
degrees Fahrenheit, 29.92 in. Hg) by using equation 5-1.
equation 5-1
![]()
Where:
K1 = 0.3869 °K/mm Hg for metric units.
= 17.71 °R/in. Hg for English units.
Equation 5-1 may be
used as written. However, if the leakage
rate observed during any of the mandatory leak checks (for example, the
post-test leak check or leak checks conducted before component changes) exceeds
La, equation 5-1 shall be modified in the following manner:
(A) Case I. No component changes made during sampling
run. In this case, replace Vm
in equation 5-1 with the following expression:
![]()
(B) Case II.
One or more component changes made during the sampling run. In this
case, replace Vm in equation 5-1 by the
following expression:
![]()
and substitute only for those leakage
rates (Li or Lp) that exceed
(iv) Volume of water vapor.
equation 5-2
![]()
Where:
(v) Moisture content.
equation 5-3
![]()
In saturated or water
droplet-laden gas streams, 2 calculations of the moisture content of the stack
gas shall be made: 1 from the impinger analysis (equation 5-3), and a second
from the assumption of saturated conditions. The lower of the 2 values of Bws
shall be considered correct. The procedure for determining the moisture content
based upon assumption of saturated conditions is given in the note of section
1.2 of method 4. For the purpose of this
method, the average stack gas temperature from figure 104 may be used to make
the determination, if the accuracy of the instack temperature sensor is ±1
degree Centigrade (2 degrees Fahrenheit).
(vi) Solvent blank concentration.
equation 5-4
![]()
(vii) Solvent wash
blank.
equation 5-5
![]()
(viii) Total
particulate weight. Determine the total particulate catch from the sum of the
weights obtained from containers 1, 1A, and 2 less the wash solvent blank (see
figure 106).
Refer to subdivision
(d)(i)(E) of this rule to assist in the calculation of
results involving 2 or more pairs of filters or 2 or more sampling trains.
(ix) Sampled gas density. Determine the density of the gas sampled from
the stack, at standard conditions (lb/ft.³).
equation 5-6
![]()
(x) Total weight of gas sampled (lbs).
equation 5-7
![]()
(xi)
Particulate concentration (lbs/1000 lbs).
equation 5-8
![]()
(xii) Excess air and
moisture correction factors:
(A) Correction factor to 50% excess air for those
sources with or without any particulate collector where no increase in moisture
content of the exhaust gas occurs after the process and before the point of
sampling.
equation 5-9
![]()
(B) Correction factor to 50% excess air for those
sources with a wet collection device (scrubber) that increases the moisture content of the exhaust gas after
the process and before the point of sampling.
equation 5-10
![]()
(C) Correction factor to convert the actual
concentration, Cs, to dry conditions.
equation 5-11
![]()
(xiii) Converted particulate concentrations, where
applicable under the department’s rules or permit.
equation 5-12
![]()
equation 5-13
![]()
equation 5-14
![]()
(xiv) Mass emission rate (lb/hr).
equation 5-15
![]()
Where:
K3 = 63.77 for English units.
(xv) Isokinetic variation:
(A) Calculation from raw data.
equation 5-16
![]()
Where:
K4 = 0.003458 mm Hg - m³ml - °K for metric
units.
=
0.002672 in. Hg - ft.³/ml - °R for English units.
(B) Calculation from intermediate values.
equation 5-17
![]()
Where:
K5 = 4.307 for metric units.
= 0.09409 for English units.
(xvi) Acceptable results. If 90%≤I≤110%,
then the results are acceptable.
If the results are low in comparison to the standard and I is beyond the
acceptable range, or if I is less than 90%, then the department may opt to
accept the results. Otherwise, reject
the results and repeat the test.
(g) Bibliography:
(i) Federal Register, Volume 42, No. 160, Part
160, Chapter 1, Title 40, Appendix A, Method 5,
(ii) Martin, Robert
M. Construction Details of Isokinetic Source Sampling Equipment. Environmental Protection
Agency.
(iii) Rom, Jerome J.
Maintenance, Calibration, and Operation of Isokinetic Source Sampling
Equipment. Environmental
Protection Agency.
(iv) Shigehara, R. T. "Adjustments in the
EPA Nomograph for Different Pitot Tube Coefficients and Dry Molecular
Weights." Stack Sampling
News, 2:4 - 11. October, 1974.
(v) Guidelines for Source Testing of Particulate. Michigan Department of
Natural Resources, Air Quality Division.
History: 1985 MR 2, Eff. Feb. 22, 1985; 1992 MR 9, Eff. Oct. 31, 1992; 2002 MR 5, Eff. Mar. 19, 2002; 2004 MR 18, Eff. Oct. 15, 2004; 2005 MR 8, Eff. April 29, 2005.
R 336.2012 Reference test method 5C.
Rule
1012. Reference test method 5C,
outstack filtration method, reads as follows:
(a) The principle, applicability, and performance
test criteria are as follows:
(i) Principle.
Particulate matter is withdrawn isokinetically from the source and
collected on solid filtering media maintained at a temperature in the range of
120 ±14 degrees Centigrade (248 ±25 degrees Fahrenheit) or such other
temperature as specified by the department’s rules or a permit condition, or as
approved by the department for a particular application. The particulate mass, which includes any
material that condenses at or above the filtration temperature, is determined
gravimetrically after removal of uncombined water.
(ii) Applicability. This method is applicable for the
determination of particulate emissions from stationary sources as identified in
table 31 of R 336.1331. The method
is also applicable when specifically provided for in the department’s rules,
orders, a permit to install, or a permit to operate.
(iii) Performance test criteria:
(A) A performance test shall consist of a minimum
of 3 separate samples of a specific air contaminant conducted within a 36-hour
period, unless otherwise authorized by the department. Each of the 3 separate samples shall be
obtained while the source is operating at a similar production level. For the purpose of determining compliance
with an applicable emission limit, rule, or permit condition, the arithmetic
mean of results of the 3 samples shall apply.
If a sample is accidentally lost or conditions occur in which 1 of the 3
samples must be discontinued because of forced shutdown, failure of an
irreplaceable portion of the sampling train, extreme meteorological conditions,
or other circumstances beyond the owner's or operator's control, compliance
may, upon the approval of the department, be determined using the arithmetic
mean of the results of 2 samples.
(B) For any source that is subject to an emission
limitation calculated to 50% excess air, the multipoint, integrated sampling
procedure of R 336.2004(1)(c) shall be used for
gas analysis. For all other sources that
require a determination of the molecular weight of the exhaust, any optional
sampling procedure of R 336.2004(1)(c) may be used. Alternatives or modifications to procedures
are subject to the approval of the department.
(C) The minimum volume per sample shall be 30
cubic feet of dry gas corrected to standard conditions (70 degrees Fahrenheit,
29.92 inches mercury). Minimum sample
time shall be 60 minutes, which may be continuous or a combination of shorter
sampling periods for sources that operate in a cyclic manner. Smaller sampling times or sample volumes,
when necessitated by process variables or other factors, may be approved by the
department.
(D) For any source whose emission control device
alters the moisture content of the exhaust gas, a moisture determination shall
be performed in a location upstream from the emission control device and in
accordance with R 336.2004(1)(d) or an
alternative method approved by the department.
(b) The following provisions apply to apparatus:
(i) Sampling train. A schematic of the sampling train used in
this method is shown in figure 103.
Construction details for many, but not all, of the train components are
given in APTD-0581 (subdivision (g)(ii) of this
rule). For changes from the APTD-0581
document and for allowable modifications to figure 103, consult with the
department.
The operating and
maintenance procedures for many, but not all, of the sampling train are
described in APTD-0576 (subdivision (g)(iii) of this
rule). Since correct usage is important
in obtaining valid results, all users shall read APTD-0576 and adopt the applicable
operating and maintenance procedures outlined in it, unless otherwise specified
herein. The sampling train consists of the following
components:
(A) Probe nozzle.
Stainless steel (316) or glass with sharp,
tapered leading edge. The angle of taper
shall be less than 30 degrees and the taper shall be on the outside to preserve
a constant internal diameter. The probe
nozzle shall be of the button-hook design, unless otherwise specified by the
department. If made of stainless steel,
the nozzle shall be constructed from seamless tubing. Other materials of construction may be used,
subject to the approval of the department.
A range of nozzle
sizes suitable for isokinetic sampling shall be available, for example, 0.32 to
1.27 cm (1/8 to 1/2 in.) - or larger if higher volume sampling trains are used
- inside diameter (ID) nozzles in increments of 0.16 cm (1/16 in.). Each nozzle shall be calibrated according to
the procedures outlined in subdivison (e) of this rule.
(B) Probe liner.
Borosilicate or quartz glass tubing with a heating system capable of
maintaining a gas temperature at the exit end during sampling of 120 ±14
degrees Centigrade (248 ±25 degrees Fahrenheit), another temperature as
specified by the department’s rules, or a temperature approved by the
department for a particular application.
The tester may opt to operate the equipment at a temperature lower than
that specified. Since the actual
temperature at the outlet of the probe is not usually monitored during
sampling, probes constructed according to APTD-0581 which utilize the calibration
curves of APTD-0576, or calibrated according to the procedure outlined in
APTD-0576, are acceptable.
Either borosilicate or
quartz glass probe liners may be used for stack temperatures up to about 480
degrees Centigrade (900 degrees Fahrenheit); quartz liners shall be used for
temperatures between 480 and 900 degrees Centigrade (900 and 1,650 degrees
Fahrenheit). Both types of liners may be
used at higher temperatures than specified for short periods of time, subject
to the approval of the department. The
softening temperature for borosilicate is 820 degrees Centigrade (1,508 degrees
Fahrenheit) and for quartz it is 1,500 degrees Centigrade (2,732 degrees
Fahrenheit).
When practical, every
effort shall be made to use borosilicate or quartz glass probe liners.
Alternatively, metal liners, such as 316 stainless steel,
Incoloy 825, or other corrosion-resistant materials made of seamless
tubing, may be used, subject to the approval of the department.
(C) Pitot tube.
Type S, as described in section 2.1 of method 2, or other device
approved by the department. The pitot tube shall be attached to the probe, as shown in
figure 103, to allow constant monitoring of the stack gas velocity. The impact (high pressure) opening plane of
the pitot tube shall be even with or above the nozzle
entry plane (see method 2, figure 2-6b) during sampling. The type S pitot
tube assembly shall have a known coefficient, determined as outlined in section
4 of method 2.
(D) Differential pressure gauge. Incline manometer or equivalent devices (2),
as described in section 2.2 of method 2.
One manometer shall be used for velocity head ( p)
readings, and the other shall be used for orifice differential pressure
readings.
(E) Filter holders. Two separate filter holders in series or 1
filter holder with separate filter supports and seals for 2 filters. One filter holder with 2 filters held in
contact with each other is not acceptable.
Materials of construction may be stainless steel (316), glass, teflon, or such other material approved by the department.
(F) Filter heating system. Any heating system capable of maintaining a
temperature around the filter holder during sampling of 120 ±14 degrees
Centigrade (248 ±25 degrees Fahrenheit), another temperature as specified by
the department’s rules or a permit condition, or a temperature approved by the
department for a particular application.
Alternatively, the tester may opt to operate the equipment at a
temperature lower than that specified. A
temperature gauge capable of measuring temperature to within 3 degrees
Centigrade (5.4 degrees Fahrenheit) shall be installed so that the temperature
around the filter holders can be regulated and monitored during sampling. Heating systems other than the one shown in
APTD-0581 may be used.
(G) Condenser.
The following system shall be used to determine the stack gas moisture
content: Three impingers connected in
series with leak-free ground glass fittings or any similar leak-free
noncontaminating fittings. All impingers shall be of the Greenburg-Smith design
and shall be modified by replacing the tip with a 1.3 cm (1/2 in.) ID glass tube extending to about 1.3 cm (1/2 in.) from the bottom
of the flask. Modifications, such
as using flexible connections between the impingers or using materials other
than glass, are permitted subject to the approval of the department’s staff.
The first impinger shall contain a known quantity of water (subdivision (d)(i)(C) of this rule), the second shall be empty, and the
third shall contain a known weight of silica gel or equivalent desiccant.
Alternatively, any
system that cools the sample gas stream and allows measurement of the water
condensed and moisture leaving the condenser, each to within 1 ml or 1 g, may
be used subject to the approval of the department. In any case, the means for measuring the
moisture leaving the condenser shall be by passing the sample gas stream
through a tared silica gel, or equivalent desiccant, trap with exit gases kept below
20 degrees Centigrade (68 degrees Fahrenheit) and determining the weight gain.
If a determination of
the particulate matter collected in the impingers is required by the
department’s rules, a permit to install, or a permit to operate, the impinger
system described in this subparagraph shall be used, without modification. Contact the department as to the sample
recovery and analysis of the impinger contents.
(H) Metering system. Vacuum gauge, leak-free pump, thermometers
capable of measuring temperature to within 3 degrees Centigrade (5.4 degrees
Fahrenheit), dry-gas meter capable of measuring volume to within 2%, and related equipment as shown in figure 103. Other metering systems capable of maintaining
sampling rates within 10% of isokinetic and capable of determining sample
volumes to within 2% may be used, subject to the approval of the department. When the
metering system is used in conjunction with a pitot tube, the system shall
enable checks of isokinetic rates.
Sampling trains
utilizing metering systems designed for higher flow rates than those described
in APTD-0581 or APTD-0576 may be used if the specifications of this method are
met.
(I) Barometer. Mercury, aneroid, or other
barometer capable of measuring atmospheric pressure to within 2.5 mm Hg (0.1
in. Hg).
In many cases, the barometric reading may be obtained from a nearby
national weather service station. When
obtained from this source, the station value, which is the absolute barometric
pressure, shall be requested and an adjustment for elevation differences between
the weather station and sampling point shall be applied at a rate of minus 2.5
mm Hg (0.1 in. Hg) per 30 m (100 ft.) elevation increase or
vice versa for elevation decrease.
(J) Gas density determination equipment. Temperature sensor and
pressure gauge, as described in sections 2.3 and 2.4 of method 2, and gas
analyzer, if necessary, as described in method 3. The temperature sensor shall, preferably, be
permanently attached to the pitot tube or sampling
probe in a fixed configuration such that the tip of the sensor extends beyond
the leading edge of the probe sheath and does not touch any metal. Alternatively, the sensor may be attached
just prior to use in the field. Note,
however, that if the temperature sensor is attached in the field, the sensor
shall be placed in an interference-free arrangement with respect to the type S
pitot tube openings (see method 2, figure 2.7).
As a second alternative, if a difference of not more than 1% in the
average velocity measurement is to be introduced, the temperature gauge need
not be attached to the probe or pitot tube. This alternative is subject to the approval
of the department.
(ii) Sample recovery. The following items are needed:
(A) Probe-liner and probe-nozzle brushes. Nylon bristle brushes with stainless steel
wire handles. The probe brush shall have
extensions, at least as long as the probe, made of
stainless steel, nylon, teflon, or similarly inert material. The brushes shall be properly sized and
shaped to brush out the probe liner and nozzle.
(B) Wash bottles--2. Glass wash bottles are recommended;
polyethylene wash bottles may be used at the option of the tester. It is recommended that acetone not be stored
in polyethylene bottles for longer than a month.
(C) Glass sample storage containers. Chemically resistant,
borosilicate glass bottles, for acetone washes, 500 ml or 1000 ml. Screw cap liners shall either be
rubber-backed teflon or shall be constructed so as to
be leak-free and resistant to chemical attack by acetone. Narrow-mouth glass bottles have been found to
be less prone to leakage. Alternatively,
polyethylene bottles may be used.
(D) Filter containers. Glass, polyethylene, or
aluminum tube containers, unless otherwise specified by the department.
(E) Graduated cylinder or balance. To measure condensed water
to within 1 ml or 1 g. Graduated
cylinders shall have subdivisions of not more than 2 ml. Most laboratory balances are capable of
weighing to the nearest 0.5 g or less.
Any of these balances are suitable for use here and in paragraph (iii)(D) of this subdivision.
(F) Plastic storage containers. Airtight containers to store silica gel.
(G) Funnel and rubber policeman. To aid in the transfer of
silica gel to container; not necessary if silica gel is weighed in the field.
(H) Funnel.
Glass or polyethylene, to aid in sample recovery.
(iii) Analysis.
The following equipment is needed for analysis:
(A) Glass weighing dishes.
(B) Desiccator.
(C) Analytical balance. To measure to within 0.1
mg.
(D) Balance.
To measure to within 0.5 mg.
(E) Beakers.
250 ml.
(F) Hygrometer.
To measure the relative humidity of the laboratory
environment.
(G) Temperature gauge. To measure the temperature
of the laboratory environment.
(c) The following provisions apply to reagents:
(i) Sampling.
The reagents used in sampling are as follows:
(A) Filters.
Two outstack filters may be any combination of alundum ceramic thimble
filters, type RA-98 or glass fiber filters, type A without organic binder. The size of such filters shall allow proper
sampling rates to maintain isokinetics using the nozzle sizes specified in
subdivision (b)(i)(A) of this rule.
Alternatively, other
types of filters may be used, subject to the approval of the department.
(B) Silica gel.
Indicating type, 6 to 16 mesh. If
previously used, dry at 175 degrees Centigrade (350 degrees Fahrenheit) for 2
hours. New silica gel may be used as received.
Alternatively, other types of desiccants (equivalent or better) may be
used, subject to the approval of the department.
(C) Water.
When analysis of the material caught in the impingers is required,
distilled water shall be used. Run
blanks prior to field use to eliminate a high blank on test samples.
(D) Crushed ice.
(E) Stopcock grease. Acetone-insoluble,
heat-stable silicone grease. This is not necessary if screw-on
connectors with teflon sleeves, or equivalent, are
used. Alternatively, other types of
stopcock grease may be used, subject to the approval of the department.
(ii) Sample recovery. Washing solvent. Either acetone or distilled water may be used
for sample recovery. If acetone is used
for washing solvent, then reagent grade, less than 0.001% residue, in glass
bottles is required. Acetone from metal
containers generally has a high residue blank and shall not be used. Suppliers sometimes transfer acetone to glass
bottles from metal containers; thus, acetone blanks shall be run prior to field
use and only acetone with low blank values (less than 0.001%) shall be used. A blank value of more than 0.001% of the
weight of acetone used shall not be
subtracted from the sample weight.
If distilled water is
used for washing solvent, use distilled water with less than 0.001%
residue. Run blanks before field use to
eliminate a high blank on test samples.
(iii) Analysis.
Two reagents are required for the analysis:
(A) Solvent. Same as paragraph (ii) of
this subdivision for quantitative transfer.
(B) Desiccant.
Anhydrous calcium sulfate, indicating type.
Alternatively, other types of desiccants may be used, subject to the approval
of the department.
(d) The following provisions apply to procedure:
(i) Sampling.
The complexity of this method is such that, in order to obtain reliable
results, testers shall be trained and experienced with the test
procedures. Sampling shall comply with
the following provisions:
(A) Pretest preparation. All the components shall be maintained and
calibrated according to the applicable procedures described in APTD-0576,
unless otherwise specified in this rule.
Weigh several 200 to
300 g portions of silica gel in airtight containers to the nearest 0.5 g.
Record the total weight of the silica gel plus container on each
container. As an alternative, the silica
gel need not be preweighed, but may be weighed directly in its impinger or
sampling holder just prior to train assembly.
Check filters visually
against light for irregularities, flaws, pinhole leaks, or cracks. Label filters of the proper size on the back
side using numbering machine ink. As an
alternative, label the shipping containers (subdivision (b)(ii)(D)
of this rule) and keep the filters in these containers at all times, except
during sampling and weighing.
Dry the filters in an
oven at 105 degrees Centigrade (220 degrees Fahrenheit) for a minimum of 2
hours, cool for at least 1 hour in a desiccator containing anhydrous calcium
sulfate, and individually weigh and record their weights to the nearest 0.1
mg. During the weighing, the filters shall
not be exposed to the laboratory atmosphere for a period of more than 2 minutes
and a relative humidity above 50%.
Procedures, other than those specified, that account for relative
humidity effects may be used, subject to the approval of the department.
(B) Preliminary determinations. Select the sampling site and the minimum
number of sampling points according to method 1 or as specified by the
department. Determine the stack
pressure, temperature, and the range of velocity heads using method 2; it is
recommended that a leak check of the pitot lines (see method 2, section 3.1) be
performed. Determine the moisture
content using approximation method 4, or its alternatives, for the purpose of
making isokinetic sampling rate settings.
Determine the stack gas dry molecular weight, as described in method 2,
section 3.6; if integrated method 3 sampling is used for molecular weight
determination, the integrated bag sample shall be taken simultaneously with,
and for the same total length of time as, the particulate sample run.
Select a nozzle size
based on the range of velocity heads so that it is not necessary to change the
nozzle size in order to maintain isokinetic sampling rates. During the run, do not change the nozzle
size. Ensure that the proper
differential pressure gauge is chosen for the range of velocity heads
encountered (see section 2.2 of method 2).
Select a suitable
probe liner and probe length so that all traverse points can be sampled. For large stacks, consider sampling from
opposite sides of the stack to reduce the length of probes.
Select a total
sampling time greater than or equal to the minimum total sampling time
specified in the test procedures for the specific industry so that the sampling
time per point is not less than 5 minutes, unless approved by the department,
or some greater time interval as specified by the department, and so that the
sample volume taken, corrected to standard conditions, exceeds the required
minimum total gas sample volume. The
latter is based on an approximate average sampling rate.
It is recommended that
the number of minutes sampled at each point be an integer or an integer plus
1/2 minute to avoid timekeeping errors.
In some circumstances,
such as in batch cycles, it may be necessary to sample for shorter times at the
traverse points and to obtain smaller gas sample volumes. In these cases, the department’s approval
shall first be obtained.
(C) Preparation of collection train. During preparation and assembly of the
sampling train, keep all openings where contamination can occur covered until
just before assembly or until sampling is about to begin.
Place 100 ml of water
in the first impinger, leave the second impinger empty, and transfer
approximately 200 to 300 g of preweighed silica gel from its container to the
third impinger. More silica gel may be
used, but care should be taken to ensure that it is not entrained and carried
out from the impinger during sampling.
Place the container in a clean place for later use in the sample
recovery. Alternatively, the weight of
the silica gel plus impinger may be determined to the nearest 0.5 g and
recorded.
Using tweezers or
clean disposable surgical gloves, place a labeled (identified) and weighed
filter in the filter holder. Be sure
that the filter is properly centered and the gasket properly placed so as to
prevent the sample gas stream from circumventing the filter. Check the filter for tears after assembly is
completed.
When glass liners are
used, install the selected nozzle using a Viton A 0ring when stack temperatures
are less than 260 degrees Centigrade (500 degrees Fahrenheit) and an asbestos
string gasket when temperatures are higher.
See APTD-0576 for details. Other
connecting systems using either 310 stainless steel or teflon
ferrules may be used. When metal liners are used, install the nozzle in the
same manner as for glass liners or by a leak-free direct mechanical
connection. Mark the probe with
heat-resistant tape or by some other method to denote the proper distance into
the stack or duct for each sampling point.
Set up the train as in
figure 103. If necessary, use a very
light coat of silicone grease on all ground glass joints. Grease only the outer portion (see APTD-0576)
to avoid the possibility of contamination by the silicone grease.
Place crushed ice around
the impingers.
(D) Leak check
procedures:
(1) Pretest leak check. A pretest leak check
is strongly recommended, but not required, to prevent invalid sampling and
wasted time. If the tester opts to
conduct the pretest leak check, the following procedure shall be used:
After the sampling
train has been assembled, turn it on and set the filter and probe heating
systems at the desired operating temperatures. Allow time for the temperatures
to stabilize. If a Viton A 0-ring or
other leak-free connection is used in assembling the probe nozzle to the probe
liner, leakcheck the train at the sampling site by plugging the nozzle and
pulling a 380 mm Hg (15 in. Hg) vacuum. A lower vacuum may be used, if it is not
exceeded during the test.
If an asbestos string
is used, do not connect the probe to the train during the leak check. Instead, leak check
the train by first plugging the inlet to the filter holder (cyclone, if
applicable) and pulling a 380 mm Hg (15 in. Hg) vacuum. A lower vacuum may be used if it is not
exceeded during the test. Then connect
the probe to the train and leak check at about 25 mm Hg (1 in. Hg) vacuum; alternatively, the probe may be leak checked with the
rest of the sampling train, in 1 step, at 380 mm Hg (15 in. Hg) vacuum. Leakage
rates in excess of 4% of the average sampling rate or 0.00057
m³/min (0.02 cfm), whichever is less, are unacceptable.
The following leak
check instructions for the sampling train described in APTD-0576 and APTD-0581
may be helpful. Start the pump with the
bypass valve fully open and the coarse adjust valve completely closed. Partially open the coarse
adjust valve and slowly close the bypass valve until the desired vacuum is
reached. Do not reverse the direction of
the bypass valve; this will cause water to back up into the filter holder. If the desired vacuum is exceeded, either
leak check at this higher vacuum or end the leak check and start over.
When the leak check is
completed, first slowly remove the plug from the inlet to the probe, filter holder, or cyclone (if applicable) and immediately turn off
the vacuum pump. This prevents the water
in the impingers from being forced backward into the filter holder and prevents
silica gel from being entrained backward into the third impinger.
(2) Leak checks during sample run. If, during the sampling run, a component
(such as a filter assembly or impinger) change becomes necessary, a leak check
shall be conducted immediately before the change is made. The leak check shall
be done according to the procedure outlined in paragraph (i)(D)(1) of this
subdivision, except that it shall be done at a vacuum equal to or greater than
the maximum value recorded up to that point in the test. If the leakage rate is found to be not more
than 0.00057 m³/min (0.02 cfm) or 4% of the average sampling rate, whichever is
less, the results are acceptable and no correction need be applied to the total
volume of dry gas metered; if, however, a higher leakage rate is obtained, the
tester shall either record the leakage rate and plan to correct the sample
volume, as shown in subdivision (f)(iii) of this rule, or shall void the
sampling run.
Immediately after
component changes, leak checks are optional; if such leak checks are done, the
procedure outlined in paragraph (i)(D)(1) of this subdivision shall be used.
(3) Post-test leak check. A leak check is mandatory at the conclusion
of each sampling run. The leak check
shall be done in accordance with the procedures outlined in paragraph (i)(D)(1)
of this subdivision, except that it shall be conducted at a vacuum equal to or
greater than the maximum value reached during the sampling run. If the leakage rate is found to be not more than 0.00057 m³/min (0.02 cfm) or 4% of the average
sampling rate, whichever is less, the results are acceptable and no correction
need be applied to the total volume of dry gas metered. If, however, a higher leakage rate is
obtained, the tester shall either record the leakage rate and correct the
sample volume, as shown in subdivision (f)(iii) of this
rule, or shall void the sampling run.
(E) Particulate train operation. During the sampling run, maintain an
isokinetic sampling rate that is within 10% of true isokinetic, unless
otherwise specified by the department.
For each run, record
the data required on a data sheet such as the one shown in figure 104. Be sure to record the initial dry-gas meter
reading. Record the dry-gas meter
readings at the beginning and end of each sampling time increment, when changes
in flow rates are made, before and after each leak check, and when sampling is
halted. Take other readings required by
figure 104 at least once at each sample point during each time increment, and
take additional readings when significant changes (20% variation in velocity
head readings) necessitate additional adjustments in flow rate. Level and zero the manometer. Because the manometer level and zero may
drift due to vibrations and temperature changes, make periodic checks during
the traverse.
Clean the portholes
prior to the test run to minimize the chance of sampling deposited
material. To begin sampling, remove the
nozzle cap and verify that the pitot tube and probe
are properly positioned. Position the
nozzle at the first traverse point with the tip pointing directly into the gas stream. Immediately start the pump and adjust the
flow to isokinetic conditions. Nomographs that aid in the rapid adjustment of
the isokinetic sampling rate without excessive computations are available. These nomographs are designed for use when
the type S pitot tube coefficient is 0.85 ±0.02 and
the stack gas equivalent density (dry molecular weight) is equal to 29 ±4. APTD-0576 details the procedure for using the
nomographs. If Cp and Md
are outside the above stated ranges, do not use the nomographs uless
appropriate steps (see subdivision (g)(iv) of this
rule) are taken to compensate for the deviations.
When the stack is
under significant negative pressure (height of impinger stem), take care to
pull low flow when inserting the probe into the stack to prevent water from
backing into the sample tubing and to avoid pulsation through the filter and
possible loss of materials.
When the probe is in
position, block off the openings around the probe and porthole to prevent
unrepresentative dilution of the gas stream.
Traverse the stack
cross section, as required by method 1 or as specified by the department, being
careful not to bump the probe nozzle into the stack walls when sampling near
the walls or when removing or inserting the probe through the portholes; this
minimizes the chance of extracting deposited material.
During the test run,
add more ice and, if necessary, salt to maintain a temperature of less than 20
degrees Centigrade (68 degrees Fahrenheit) at the condenser/silica gel
outlet. Also, periodically check the
level and zero of the manometer.
If the pressure drop
across the filter becomes too high and makes isokinetic sampling difficult to
maintain, the filter may be replaced in the midst of a sample run. It is recommended that another complete
filter assembly be used rather than attempting to change the filter
itself. Before a new filter assembly is
installed, conduct a leak check (see paragraph (i)(D)(2)
of this subdivision). The total
particulate weight shall include the summation of all filter assembly catches.
A single train shall
be used for the entire sample run, except in cases where simultaneous sampling
is required in 2 or more separate ducts, at 2 or more different locations
within the same duct, or where equipment failure necessitates a change of
trains. In all other situations, the use
of 2 or more trains shall be subject to the approval of the department.
Note that when 2 or
more trains are used, separate analyses of the front-half catches from the
individual trains may be combined, as may the impinger catches, and 1 analysis
of the front-half catch and 1 analysis of impinger catch may be performed. Consult with the department for details
concerning the calculation of results when 2 or more trains are used.
At the end of the
sample run, turn off the coarse adjust valve, remove the probe and nozzle from
the stack, turn off the pump, record the final dry-gas meter reading, and
conduct a post-test leak check, as outlined in paragraph (i)(D)(3). Also, leak check the pitot lines as described
in method 2, section 3.1; the lines shall pass this leak check to validate the
velocity head data.
(F) Calculation of percent isokinetic. Calculate percent isokinetic (see subdivision
(f) of this rule) to determine whether the run was valid or whether another test
run should be
made. If there
was difficulty in maintaining
isokinetic rates due to source conditions, consult with the department for
possible variance on the isokinetic rates.
(ii) Sample recovery. Proper cleanup procedure begins as soon as
the probe is removed from the stack at the end of the sampling period. Allow
the probe to cool.
When the probe can be
safely handled, wipe off all external particulate matter near the tip of the
probe nozzle and place a cap over it to prevent losing or gaining particulate
matter. Do not cap off the probe tip
tightly while the sampling train is cooling down as this creates a vacuum in
the filter holder and draws water from the impingers into the filter holder.
Before moving the sample
train to the cleanup site, remove the probe from the sample train, wipe off the
silicone grease, and cap the open outlet of the probe. Be careful not to lose any condensate that
might be present. Wipe off the silicone grease from the filter inlet where the
probe was fastened and cap it. Remove
the umbilical cord from the last impinger and cap the impinger. If a flexible line is used between the first
impinger or condenser and the filter holder, disconnect the line at the filter
holder and let any condensed water or liquid drain into the impingers or
condenser. After wiping off the silicone
grease, cap off the filter holder outlet and impinger inlet. Ground-glass
stoppers, plastic caps, or serum caps may be used to close these openings.
Transfer the probe
and filter-impinger assembly to the cleanup area. This area shall be clean and protected from
the wind so that the chances of contaminating or losing the sample are
minimized.
Save a portion of the
solvent used for cleanup as a blank.
Take 200 ml of this solvent directly from the wash bottle being used and
place it in a glass sample container labeled "solvent blank."
Inspect the train
prior to and during disassembly and note any abnormal conditions. Treat the samples as follows:
Container
Nos. 1, 1A. Carefully remove the
filters from the filter holders and place in their identified containers. Use a pair of tweezers or clean disposable
surgical gloves, or both, to handle the filters. Carefully transfer to the container any
particulate matter or filter fibers, or both, that adhere to the filter holder
gasket by using a dry nylon bristle brush or sharp-edged blade, or both. Seal the container.
Container
No. 2. Taking care to see that
dust on the outside of the probe or other exterior surfaces does not get into
the sample, the testor shall quantitatively recover from particulate matter or
any condensate from the nozzle, probe fitting, probe liner, and from both
filter holders by washing these components with solvent and placing the wash in
a glass container.
Perform the solvent
rinses as follows:
Carefully remove the
probe nozzle and clean the inside surface by rinsing with solvent from a wash
bottle and brushing with a nylon bristle brush.
Brush until the solvent rinse shows no visible particles and then make a
final rinse of the inside surface with solvent.
Brush and rinse the
inside parts of the Swagelok fitting with solvent in a similar way until no
visible particles remain.
Rinse the probe liner
with solvent by tilting and rotating the probe while squirting solvent into its
upper end so that all inside surfaces are wetted with acetone. Let the solvent drain from the lower end into
the sample container. A glass or
polyethylene funnel may be used to aid in transferring liquid washes to the
container. Follow the solvent rinse with
a probe brush. Hold the probe in an
inclined position and squirt solvent into the upper end as the probe brush is
being pushed with a twisting action through the probe; hold a sample container
underneath the lower end of the probe and catch any solvent and particulate
matter that is brushed from the probe.
Run the brush through the probe 3 or more times until no visible
particulate matter is carried out with the solvent or until none remains in the
probe liner on visual inspection. With
stainless steel or other metal probes, run the brush through, in the above
prescribed manner, not less than 6 times, since metal probes have small
crevices in which particulate matter can be entrapped. Rinse the brush with solvent and
quantitatively collect these washings in the sample container. After the brushing, make a final solvent
rinse of the probe as described above.
It is recommended
that 2 people be used to clean the probe to minimize sample losses. Between sampling runs, keep brushes clean and
protected from contamination.
After ensuring that
all joints have been wiped clean of silicone grease, clean the inside of both
filter holders by rubbing the surfaces with a nylon bristle brush and rinsing
with solvent. Rinse each surface 3
times, or more if needed, to remove visible
particulate. Make a final rinse of the
brush and filter holder.
After all solvent
washings and particulate matter have been collected in the sample container, tighten the lid on the sample container so that
solvent will not leak out when it is shipped to the laboratory. Mark the height of the fluid level to
determine whether or not leakage occurred during transport. Label the container to clearly identify its
contents.
Container
No. 3. Note the color of the
indicating silica gel to determine if it has been completely spent and make a
notation of its condition. Transfer the
silica gel from the third impinger to its original container and seal. A funnel may make it easier to pour the
silica gel without spilling it. A rubber
policeman may be used as an aid in removing the silica gel from the
impinger. It is not necessary to remove
the small amount of dust particles that adhere to the impinger wall and are
difficult to remove. Since the gain in
weight is to be used for moisture calculations, do not use any water or other
liquids to transfer the silica gel. If a
balance is available in the field, follow the procedure for container No. 3 in
paragraph (iii) of this subdivision.
Impinger
water. Treat the impingers as
follows: Make a notation of any color or
film in the liquid catch. Measure the
liquid that is in the first 2 impingers to within ±1 ml by using a graduated
cylinder or by weighing it to within ±1.0 g by using a balance if none is
available. Record the volume or weight of liquid present. This information is required to calculate the
moisture content of the effluent gas.
Discard the liquid
after measuring and recording the volume or weight, unless analysis of the
impinger catch is required (see subdivision (b)(i)(G)
of this rule).
If a different type
of condenser is used, measure the amount of moisture condensed either
volumetrically or gravimetrically.
Whenever possible,
containers shall be shipped in a manner that keeps them upright at all times.
(iii) Analysis.
Record the data required on a sheet such as the one shown in figure
106. Handle each sample container as
follows:
Container
Nos. 1, 1A. Analyze and report
each filter separately. Transfer the filter and any loose particulate from the
sample container to a tared-glass weighing dish. Dry the filter in an oven at 105 degrees
Centigrade (220 degrees Fahrenheit) for a minimum of 2 hours, cool for at least
1 hour in a desiccator containing anhydrous calcium sulfate, and weigh and
record its weight to the nearest 0.1 mg.
During the weighing, the filter shall not be exposed to the laboratory
atmosphere for a period of more than 2 minutes or a relative humidity above
50%. Procedures, other than those
specified, that account for relative humidity effects may be used, subject to
the approval of the department. The
method used for the drying and weighing of filters shall be consistent before
and after the test.
Container
No. 2. Note the level of liquid
in the container and confirm on the analysis sheet whether or not leakage
occurred during transport. If a
noticeable amount of leakage has occurred, either void the sample or use
methods, subject to the approval of the department, to correct the final
results.
Measure the liquid in
this container either volumetrically to ±1 ml or gravimetrically to ±1.0
g. Transfer the contents to a tared
250-ml beaker and evaporate to dryness either at ambient temperature and
pressure for acetone or at 95 degrees Centigrade (203 degrees Fahrenheit) in an
oven for distilled water. Then subject
the sample to 250 degrees Centigrade (482 degrees Fahrenheit) in an oven for 2
to 3 hours. Desiccate for 24 hours and
weigh to a constant weight. Report the
results to the nearest 0.1 mg.
Container
No. 3. Weigh the spent silica
gel, or silica gel plus impinger, to the nearest 0.5 g using a balance. This step may be conducted in the field.
"Solvent
blank" container. Measure
solvent in this container either volumetrically or gravimetrically. Transfer the contents to a tared 250-ml
beaker and evaporate to dryness either at ambient temperature and pressure for
acetone or at 95 degrees Centigrade (203 degrees Fahrenheit) in an oven for
distilled water. Then subject the sample
to 250 degrees Centigrade (482 degrees Fahrenheit) in an oven for 2 to 3
hours. Desiccate for 24 hours and weigh
to a constant weight. Report the results
to the nearest 0.1 mg.
If acetone is used,
the contents of container No. 2, as well as the acetone blank container, may be
evaporated at temperatures higher than ambient.
If evaporation is done at an elevated temperature, the temperature shall
be closely supervised, and the contents of the beaker shall be swirled
occasionally to maintain an even temperature.
Use extreme care, as acetone is highly flammable and has a low flash
point.
(e) Calibration.
Maintain a laboratory log of all calibrations. Calibrations shall
comply with all of the following
provisions:
(i) Probe nozzle.
Probe nozzles shall be calibrated before their initial use in the
field. Using a micrometer, measure the
inside diameter of the nozzle to the nearest 0.025 mm (0.001 in.). Make 3 separate measurements using different diameters
each time and obtain the average of the measurements. The difference between the high and low
numbers shall not exceed 0.1 mm (0.004 in.).
When nozzles become nicked, dented, or corroded, they shall be reshaped,
sharpened, and recalibrated before use.
Each nozzle shall be permanently and uniquely identified.
(ii) Pitot tube.
The type S pitot tube assembly shall be
calibrated according to the procedure outlined in section 4 of method 2.
(iii) Metering system. Before its initial use in the field, the
metering system shall be calibrated according to the procedure outlined in
APTD-0576. Instead of physically
adjusting the dry-gas meter dial readings to correspond to the wet-test meter
readings, calibration factors may be used to mathematically correct the gas
meter dial readings to the proper values.
Before calibrating the metering system, it is suggested that a leak
check be conducted. For metering systems
having diaphragm or rotary pumps, the normal leak check procedure will not
detect leakages within the pump. For these cases, the following leak check
procedure is suggested: Make a 10-minute
calibration run at 0.00057 m³/min (0.02 cfm); at the end of the run, take the
difference of the measured wet-test meter and dry-gas meter volumes; divide the
difference by 10 to get the leak rate.
The leak rate shall not exceed 0.00057 m³/min (0.02
cfm).
After each field use,
the calibration of the metering system shall be checked by performing 3
calibration runs at a single, intermediate orifice setting, based on the
previous field test, with the vacuum set at the maximum value reached during
the test series. To adjust the vacuum,
insert a valve between the wet-test meter and the inlet of the metering
system. Calculate the average value of
the calibration factor. If the
calibration has changed by more than 5%, recalibrate the meter over the full
range of orifice settings, as outlined in APTD‑0576.
Alternatively, a
spirometer may be substituted for a wet-test meter in the above mentioned
calibration procedures.
Alternative
procedures, such as using the orifice meter coefficients, may be used, subject
to the approval of the department.
If the dry-gas meter
coefficient values obtained before and after a test series differ by more than
5%, the test series shall be performed using whichever meter coefficient value
(before or after) gives the lower value of total sample volume.
(iv) Probe heater calibration. The probe heating system shall be calibrated
before its initial use in the field according to the procedures outlined in
APTD-0576. Probes constructed according
to APTD-0581 need not be calibrated if the calibration curves in APTD-0576 are
used.
(v) Temperature gauges. Use the procedure in section 4.3 of method 2
to calibrate in-stack temperature gauges.
Dial thermometers, such as those used for the dry-gas meter and condenser
outlet, shall be calibrated against mercury-in-glass thermometers.
(vi)
Leak check of metering system shown in figure 103. That portion of the
sampling train from the pump to the orifice meter shall be leak checked before
initial use and after each shipment.
Leakage after the pump will result in less volume being recorded than is
actually sampled. The following procedure is suggested (see figure 107): Close the main valve on the meter box. Insert a 1-hole rubber stopper with rubber
tubing attached into the orifice exhaust pipe.
Disconnect and vent the low side of the orifice manometer. Close off the low side orifice tap.
Pressurize the system to 13 to 18 cm (5 to 7 in.) water column by blowing into
the rubber tubing. Pinch off the tubing
and observe the manometer for 1 minute.
A loss of pressure on the manometer indicates a leak in the meter box;
leaks, if present, shall be corrected.
(vii) Barometer.
Calibrate against a mercury barometer.
(f) Calculations.
When carrying out calculations, retain at least 1 extra decimal figure
beyond that of the acquired data. Round
off figures after the final calculation.
Other forms of the equations may be used if they give equivalent
results. All of the following provisions
apply to calculations:
(i) Nomenclature:
An = Cross-sectional area of nozzle, m² (ft ²)
A = Cross-sectional
area of stack or flue at the point of sampling, ft.².
Bws = Water
vapor in the gas stream, proportion by volume, expressed as a fraction.
Bwi = Percent
water vapor in gas entering source particulate control device, determined by
method 4.
Bwo = Percent
water vapor in gas exiting source particulate control device.
Ca =
Cs = Concentration
of particulate matter in stack gas, pounds per 1,000 pounds of actual stack
gas.
CsD = Concentration
of particulate matter in stack gas, moisture excluded, pounds per 1,000 pounds
of dry stack gas.
Cs50 = Concentration
of particulate matter corrected to 50% excess air, pounds per 1,000 pounds of
stack gas.
Cs50D = Concentration
of particulate matter corrected to 50% excess air, excluding any water addition
from a collector, pounds per 1,000 pounds of stack gas.
E = Mass
emission rate of particulate, lb/hr.
F50 = Concentration
conversion factor to 50% excess air with no moisture alterations in exhaust.
F50D = Concentration
conversion factor to 50% excess air, excluding any moisture added to exhaust
gas by pollution collection system.
FD = Concentration
conversion factor to dry basis, excluding any water in the stack gas.
I = Percent
of isokinetic sampling.
La = Maximum
acceptable leakage rate for either a pretest leak check or for a leak check
following a component change; equal to 0.00057 m³/min (0.02 cfm) or 4% of the
average sampling rate, whichever is less.
Li = Individual
leakage rate observed during the leak check conducted before the
"ith" component change (i = 1, 2, 3 . . . . . n), m³/min (cfm).
Lp = Leakage rate observed during the post-test
leak check, m³/min (cfm).
Md = Molecular
weight of dry stack gas, g/g mole (lb/lb-mole), calculated by method 3,
equation 3-2, using data from integrated method 3.
mn = Total amount of particulate matter collected,
mg.
Mw = Molecular
weight of water, 18.0 g/g-mole (18.0 lb/lb-mole).
ma = Mass of residue of solvent after evaporation,
mg.
mg = Total weight of gas samples through nozzle,
lb.
Pbar = Barometric
pressure at the sampling site, mm Hg (in. Hg).
Ps = Absolute
stack gas pressure.
Pstd
= Standard absolute pressure, 760 mm Hg (29.92 in. Hg).
R = Ideal
gas constant, 0.06236 mm Hg-m³/°K-g-mole (21.85 in. Hg-ft³/°R-lb-mole).
Tm = Absolute
average dry-gas meter temperature (see figure 104), °K (°R).
Ts = Absolute
average stack gas temperature (see figure 104), °K (°R).
Tstd = Standard
absolute temperature, 294.1 °K (530 °R).
Va